Philips Medical Systems (Cleveland) Inc: Brilliance 10 Slice (Air), Model 728251 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories

Recall number
Z-0547-2019
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance 10 Slice (Air), Model 728251 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
Status
Terminated
Initiated
2018-10-18
Posted
not on file
Terminated
2021-08-03
Reason
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Root cause
Process control
Action
On October 18, 2018, Philips notices to their customers via courier service. Customers were advised to take the following actions: If any unusual noise is heard, discontinue use of the system immediately and contact your Philips representative for support. Field Order Change (FCO) 72800703 will be deployed to initiate an inspection of the front gantry cover. If the inspection finds evidence of weakening at this connection point, a temporary repair solution will be put in place to secure the cover for continued system use. FCO 72800704 will then be deployed to replace the affected front cover. A Philips Field Service Engineer will contact you to schedule the inspection of your system. Customers in need of further information or support concerning this issue, please contact your local Philips representative (North America & Canada) 1-800-722-9377.
Quantity
not on file
Distribution
US Nationwide Distribution
Lot, serial or model codes
System Serial Number: 2598 3008 3098 4001 4002 4003 4005 4006 4007 4008 4009 4010 4011 4012 4014 4015 4016 4020 4021 4022 4023 4026 4027 4028 4029 4030 4031 4032 4033 4034 4035 4039 4040 4044 4045 4047 4048 4049 4050 4052 4053 4054 4056 4058 4059 4061 4062 4063 4064 4065 4066 4067 30099 40004 40005 40006 40007 40008 40009 40010 40011 40012 40014 40015 40016 40018 40019 40020 40021 40023 40025 40028 40029 40031 40032 40034 62986554 4008/6086 4039/7128 728251_859-1766
510(k) numbers
["K012009"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register