Skip to the content

FDA Medical Device Recalls (openFDA)

Arrow International Inc: Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.

Recall number
Z-0546-2014
Recalling firm
Arrow International Inc
Product
Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
Status
Terminated
Initiated
2013-10-09
Posted
2013-12-19
Terminated
2015-09-21
Reason
A labeling inconsistency was discovered in which the lidstock states that the kit contains a stimulator cable with 1.5mm socket, however, the kit contains a stimulator cable with 1.5mm pin.
Root cause
Labeling mix-ups
Action
The firm, Arrow International, Inc., sent an "Urgent Medical Device Recall Notification" letter dated October 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any of listed affected products/lots you have in stock; return affected products, customer service representative will contact you with RGA# and instructions for the return of the product, and complete enclosed Recall Acknowledgement form and fax to 1-800-343-2935, Attn: Customer Service, even if you have not affected product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.
Quantity
121
Distribution
Worldwide distribution: US (nationwide) including states of: MA, LA, CA, GA, KS, TX, PA, NY, MO, WI, NJ, MS, CT, ID, AL, CO, and PR; and Singapore.
Lot, serial or model codes
Product number AB-07200 Stimulator Adapter Cable with lot numbers: RF0059570, RF2069779, RF0088984, RF2108983, RF0128588, RF3052266, RF1071764.
510(k) numbers
["K121403"]
PMA numbers
not on file
Product code
CAZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register