FDA Medical Device Recalls (openFDA)
Arrow International Inc: Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
- Recall number
- Z-0546-2014
- Recalling firm
- Arrow International Inc
- Product
- Continuous Peripheral Nerve Block Catheter Kit and Set Continuous Peripheral Nerve Block Catheter Kit and Set permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia or analgesia techniques for periods not exceeding 72 hours.
- Status
- Terminated
- Initiated
- 2013-10-09
- Posted
- 2013-12-19
- Terminated
- 2015-09-21
- Reason
- A labeling inconsistency was discovered in which the lidstock states that the kit contains a stimulator cable with 1.5mm socket, however, the kit contains a stimulator cable with 1.5mm pin.
- Root cause
- Labeling mix-ups
- Action
- The firm, Arrow International, Inc., sent an "Urgent Medical Device Recall Notification" letter dated October 14, 2013 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any of listed affected products/lots you have in stock; return affected products, customer service representative will contact you with RGA# and instructions for the return of the product, and complete enclosed Recall Acknowledgement form and fax to 1-800-343-2935, Attn: Customer Service, even if you have not affected product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-800-343-2935.
- Quantity
- 121
- Distribution
- Worldwide distribution: US (nationwide) including states of: MA, LA, CA, GA, KS, TX, PA, NY, MO, WI, NJ, MS, CT, ID, AL, CO, and PR; and Singapore.
- Lot, serial or model codes
- Product number AB-07200 Stimulator Adapter Cable with lot numbers: RF0059570, RF2069779, RF0088984, RF2108983, RF0128588, RF3052266, RF1071764.
- 510(k) numbers
- ["K121403"]
- PMA numbers
- not on file
- Product code
- CAZ
- Firm FEI
- 3015859709
- Firm city
- Reading
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28