Skip to the content

FDA Medical Device Recalls (openFDA)

Datex-Ohmeda Inc: Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)

Recall number
Z-0545-04
Recalling firm
Datex-Ohmeda Inc
Product
Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)
Status
Terminated
Initiated
2004-02-03
Posted
2004-07-20
Terminated
2005-11-17
Reason
Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
Root cause
Other
Action
Recall letters dated February 03, 2004 stated that a firm representative would contact the consignee to schedule a visit to replace software and hardware to correct the problem.
Quantity
1617 units in the United States
Distribution
The product was distributed nationwide to hospitals in the Unites States.
Lot, serial or model codes
The affected units in the United States have serial numbers between 40013635 and 40162379.
510(k) numbers
["K973985"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2183066
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register