Skip to the content

FDA Medical Device Recalls (openFDA)

SpineFrontier, Inc.: PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.

Recall number
Z-0544-2014
Recalling firm
SpineFrontier, Inc.
Product
PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.
Status
Terminated
Initiated
2013-10-18
Posted
2013-12-19
Terminated
2016-08-22
Reason
During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.
Root cause
Software design (manufacturing process)
Action
SpineFrontier sent an Voluntary Recall Notification dated November 5,2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. All affected implants are to be returned to SpineFrontier for inspection. SpineFrontier is requiring confirmation that you have received this notice and that you will initiate the immediate return of the implants listed above. IMMEDIATE RESPONSE REQUIRED. Please feel free to contact us at 978-232-3990 Monday through Friday 8:30AM to 5:30PM EST with any questions
Quantity
35 Devices
Distribution
US Distribution to VA and TX.
Lot, serial or model codes
Lot/Serial Numbers: 357305-000
510(k) numbers
["K123164"]
PMA numbers
not on file
Product code
MNI
Firm FEI
3005977257
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register