Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.

Recall number
Z-0542-2009
Recalling firm
Roche Diagnostics Corp.
Product
Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.
Status
Terminated
Initiated
2008-09-15
Posted
2009-01-06
Terminated
2009-12-01
Reason
Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Root cause
Package design/selection
Action
Consignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
Quantity
7909
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot numbers: 60258801, 60258601, 60145701, 60145601, 69921901, 69921801, 69786201, 69922001, 69786101, 69676101, 69509701, 69431901, 69357601, 69342001, 69246501, 68966601 and 68871601.
510(k) numbers
["K060373"]
PMA numbers
not on file
Product code
JIX
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register