FDA Medical Device Recalls (openFDA)
Roche Diagnostics Corp.: Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.
- Recall number
- Z-0542-2009
- Recalling firm
- Roche Diagnostics Corp.
- Product
- Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.
- Status
- Terminated
- Initiated
- 2008-09-15
- Posted
- 2009-01-06
- Terminated
- 2009-12-01
- Reason
- Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
- Root cause
- Package design/selection
- Action
- Consignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.
- Quantity
- 7909
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Lot numbers: 60258801, 60258601, 60145701, 60145601, 69921901, 69921801, 69786201, 69922001, 69786101, 69676101, 69509701, 69431901, 69357601, 69342001, 69246501, 68966601 and 68871601.
- 510(k) numbers
- ["K060373"]
- PMA numbers
- not on file
- Product code
- JIX
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28