FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels using three lasers (488 nm, 638 nm and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.
- Recall number
- Z-0541-2026
- Recalling firm
- Beckman Coulter Inc.
- Product
- DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels using three lasers (488 nm, 638 nm and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.
- Status
- Open, Classified
- Initiated
- 2025-09-22
- Posted
- 2025-11-18
- Terminated
- not on file
- Reason
- Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
- Root cause
- Under Investigation by firm
- Action
- On 09/22/2025. an Urgent Medical Device Recall letter was sent via postal mail and/or email informing customers that Beckman Coulter has determined that the DxFLEX software version 2.3.4.37 may crash due to a code error issue.While using the DxFLEX Flow Cytometer instrument for data acquisition using user defined acquisition protocol, the CytExpert for DxFLEX software (version 2.3.4.37) may unexpectedly crash. Due to that, the data acquisition may not be completed according to the preset conditions (i.e., number of events and/or acquisition time). When the software crash is not detected promptly, the sample may be completely consumed. Customers are instructed to: Follow the steps below when running experiments in the instrument until the software is upgraded to the version that will fix the issue: a. Prepare the experiment file (create sample, set location, etc.) b. Switch to first Tube sheet for each task to be acquired c. Save and close the experiment file d. Open the experiment again (DO NOT switch to REPORT sheet after opening experiment) This step will prevent the software from crashing due to this code error issue. e. Perform acquisition f. Analyze data. " Beckman Coulter recommends posting this letter on or near the affected analyzers until the update to the software that addresses this issue is completed. For questions: contact our Customer Support Center " From our website: http://www.beckman.com " By phone: call (800) 369-0333 in the United States.
- Quantity
- 25 units
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, CT, FL, IL, NY, OH, OR, TN, and TX. The country of India.
- Lot, serial or model codes
- Catalog Number: C78500 UDI-DI code: 15099590754013 Serial Numbers: BH44060, BJ09004, BJ14015, BJ09010, BJ13012, BJ14014, BJ13011, BJ14017, BJ13013, BJ09003, BJ09008, BJ09009, BH39057, BH44059, BJ14016, BJ05002, BJ17020, BJ17021, BH26031, BJ19024, BJ09005, BJ09006, BJ09007, BJ19022, BJ19023
- 510(k) numbers
- ["K232600"]
- PMA numbers
- not on file
- Product code
- OYE
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28