Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Giraffe Incubator infant warmer

Recall number
Z-0541-2020
Recalling firm
GE Healthcare, LLC
Product
Giraffe Incubator infant warmer
Status
Terminated
Initiated
2019-10-11
Posted
2019-12-03
Terminated
2024-08-15
Reason
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Root cause
Device Design
Action
The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
Quantity
19425 units
Distribution
worldwide
Lot, serial or model codes
Model Numbers: 6651-XXXX-XXX 2074921-00X-XXXXXX M1050206 M1050207 M1136120 M1136222 M1136226 M1138367 M1145526 M1145527 M1146091 M1146092 M1148634 M1161351 M1161353 M1161357 M1206149 M1235958 Updated 6/3/2021 - All model numbers and units are affected.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FMT
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register