Biosense Webster, Inc.: Biosense Webster PREFACE Guiding Sheath, Anterior, for the introduction of intravascular electrophysiology catheters into any cardiac chamber. Catalog Number 301803A

Recall number
Z-0540-2007
Recalling firm
Biosense Webster, Inc.
Product
Biosense Webster PREFACE Guiding Sheath, Anterior, for the introduction of intravascular electrophysiology catheters into any cardiac chamber. Catalog Number 301803A
Status
Terminated
Initiated
2006-11-17
Posted
2007-02-28
Terminated
2009-06-25
Reason
Biosense Webster, Inc. has received a number of complaints regarding the radiopaque tip separating or partially detaching from the PREFACE Sheath during use in the left atrium.
Root cause
Other
Action
Initially a Medical Device Correction Advisory letter was mailed on November 20, 2006 to all domestic and international consignees recommending to customers that the Preface Sheath was not to be used with the NaviStar ThermoCool Catheter or any 8 French or larger catheter. On December 9, 2006, a Recall Customer letter and Certification form were sent out to all customers and Johnson & Johnson affiliates who had purchased the product in the past three years. For customers in the US, two letters and forms were sent out -one addressed to the EP/Cardiology Lab and the second to the Risk Management Office. For the US, customers will return product directly to Biosense Webster afler receiving a return authorization number. Returned product will be segregated and labeled as quarantined product. Returned product will be used as necessary for the root cause investigation. Any product not used at the conclusion of the investigation will be destroyed. For international customers, product will be returned to the Biosense Webster warehouse in Roden, Netherlands or to the Biosense Webster US warehouse (depending on order placement) if export issues don''t present a problem. If product cannot be easily returned to Biosense Webster, the customers will be instructed to destroy the product and provide notification that the destruction was performed.
Quantity
182
Distribution
Worldwide distribution, including USA, Australia, Brazil, Canada, China, Colombia, Hong Kong, Indonesia, Israel, India, Japan, Korea, the Netherlands, New Zealand, Pakistan, Singapore, Thailand, and Taiwan.
Lot, serial or model codes
Lot numbers: L0103107, L0103244, L0206088, L0306027, L0306082, L0405345, L0704269, L0804090, L0902372, L1002091 & L1004055
510(k) numbers
["K001139"]
PMA numbers
not on file
Product code
DYB
Firm FEI
3003113629
Firm city
Irwindale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register