FDA Medical Device Recalls (openFDA)
ResMed Ltd.: AirFit N20 Nasal Mask and User Guide
- Recall number
- Z-0539-2024
- Recalling firm
- ResMed Ltd.
- Product
- AirFit N20 Nasal Mask and User Guide
- Status
- Open, Classified
- Initiated
- 2023-11-20
- Posted
- 2023-12-21
- Terminated
- not on file
- Reason
- Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
- Root cause
- Device Design
- Action
- On 11/20/23, "Urgent Field Safety Notices" and FAQs were placed on the firms website at www.resmed.com/magnetupdate, and these notices were emailed to healthcare providers, and distributors who were asked to do the following: 1) Provide a copy of this notice and the physician letter to prescribing physicians, and/or any other relevant healthcare professionals, notifying them of the updated labeling (contraindications and warning). 2) Provide a copy of the patient letter to all patients currently using an affected mask with magnets, notifying them of the updated labeling (contraindications and warning). Note, patients are instructed in the patient letter to contact their mask provider for a replacement mask if they are now contraindicated. 3) Provide a replacement mask for contraindicated patients to an alternative mask without magnets in a timely manner. Where an alternative mask is not available, inform patients to consult with their physician. 4) Instruct patients to consult their physician and/or manufacturer of their implant / other medical device if they require additional information on the potential adverse effects of magnetic fields for their particular device (as described in warnings). 5) For alternative mask options, please contact your ResMed or Customer Service representative. 6) Complete and return the customer acknowledgement form via email to magnetresponse@resmed.com For any questions, please contact your local ResMed contact or visit www.resmed.com/contact.
- Quantity
- 7,724,968
- Distribution
- US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA
- Lot, serial or model codes
- All mask lots used with User Guide 638218/version 2020-02, 638243/version 2020-06, and prior. UDI-DI/GTIN: 619498635016, 619498635054, 619498635061, 619498635351, 619498635047, 619498635030, 619498635078, 619498635382, 619498635023, 619498635344, 619498635009, 619498635399, 619498635115, 619498635122, 619498635375, 619498635214, 619498635221, 619498635252, 619498635245, 619498635313, 619498635337, 619498635207, 619498635320, 619498635238, 619498635092, 619498635108, 619498635306, 619498635290, 619498635184, 619498635146, 619498635160, 619498635139, 619498635177, 619498635153
- 510(k) numbers
- ["K171212"]
- PMA numbers
- not on file
- Product code
- BZD
- Firm FEI
- 3004604967
- Firm city
- Bella Vista
- Firm state
- not on file
- Firm country
- Australia
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28