Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta VBH Head FIX The VBH HeadFIX is intended for positioning and immobilization of the head and neck, stereotactic diagnostic localization and stereotactic radiotherapy of cranial targets.

Recall number
Z-0538-2011
Recalling firm
Elekta, Inc.
Product
Elekta VBH Head FIX The VBH HeadFIX is intended for positioning and immobilization of the head and neck, stereotactic diagnostic localization and stereotactic radiotherapy of cranial targets.
Status
Terminated
Initiated
2010-10-14
Posted
2010-12-03
Terminated
2010-12-28
Reason
The HeadFIX Baseplate is not screwed down to the adapter to allow to compensate for roll and pitch with the HeadFIX leveling screws of the HeadFIX Baseplate.
Root cause
Other
Action
Elekta, Inc. sent an Important Notice A343 dated October 14, 2010 to all affected consignees. The Notice identified the product, the problem, and the action to be taken by the consignees. Consignees were instructed constantly monitor the patient before and during the treatment. Check the patient's position after every patient movement (e.g. coughing). Advise all appropriate personnel of the content of the notice place a copy of the Notice In the Important Notice Section of the HeadFix User Manual P10103-108. For questions regarding this recall service.schwabmuenchen@elekta.com
Quantity
34 units
Distribution
Worldwide Distribution - USA including AZ, CA, CT, FL, GA, IL, KY, MI, MS, NY, NC, OK, OR, RI, TN, TX, VA, WA and the country of Canada
Lot, serial or model codes
P10103
510(k) numbers
["K030439"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register