Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece Component: Not a component

Recall number
Z-0537-2026
Recalling firm
Exactech, Inc.
Product
Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece Component: Not a component
Status
Open, Classified
Initiated
2025-09-05
Posted
2025-11-17
Terminated
not on file
Reason
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Root cause
Release of Material/Component prior to receiving test results
Action
On 09/05/2025, the firm emailed an "URGENT MEDICAL DEVICE RECALL NOTICE" to customers informing that, during routine inventory review, it was discovered that the a number of parts/products, which were pending design verification testing were incorrectly placed in active internal inventory rather than designated quarantine locations. Several part/products were distributed to customers. Customers are instructed to: "Carefully review this notification to ensure that they fully understand the issue and all actions required. "Immediately identify, quarantine, and return any of the subject devices in their inventory and/or customer s inventory. "Stop and cease use of any affected devices. "Notify all their accounts that may have received these products and extend this recall notice, including the Description of Issue and Potential Clinical Impact, to them. "Complete and Return the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com. For questions or assistance - contact Recall Support Team at recalls@exac.com
Quantity
1
Distribution
U.S.: SC O.U.S.: N/A
Lot, serial or model codes
Lot Code: Part Number: 301-11-11, UDI: 10885862592934, Lot Number: 410413-001
510(k) numbers
["K042021"]
PMA numbers
not on file
Product code
PHX
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register