FDA Medical Device Recalls (openFDA)
Bard Access Systems Inc.: PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
- Recall number
- Z-0536-2018
- Recalling firm
- Bard Access Systems Inc.
- Product
- PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
- Status
- Terminated
- Initiated
- 2017-12-28
- Posted
- 2018-02-06
- Terminated
- 2021-06-14
- Reason
- Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
- Root cause
- Labeling design
- Action
- Customers were notified via letter on approximately 12/28/2017. Instructions include a reminder that if additional extension sets are attached to the luers of the catheters, they should be appropriately secured as per hospital recommended securement protocol, to refrain from having unsupported extension sets that could put stress on the luer extension leg junction of the catheter, and to notify customers if the product was further distributed. Also, the notification contained a statement that Bard Access is in the process of updating their Instructions for Use to include a statement regarding the importance of properly securing additional extension sets per hospital recommended securement protocol. On 04/11/2018, the recall was expanded and customers were notified via letter, which explained that the previous recall has been expanded to involve return of affected product as well as include product distributed from February 2017 through March 2018. Instructions included to examine inventory for affected product, quarantine and immediately discontinue distribution of the product, complete and return the Business Recall Response Form, arrange for the return of affected product, and notify customers if the product has been further distributed. For distributors, Bard Access Systems requested a customer list so that they could notify their customers.
- Quantity
- 53,699 units total
- Distribution
- Distribution US nationwide.
- Lot, serial or model codes
- Product Code (Lot/Serial No.): S1274108D (REBN2334), S9274108D (REBN2336), S3274108PD (REBP1471), S3274335 (REBP1580), S3274355 (REBP1581), S3274355P (REBP1582), S3274335P (REBP1583), S1274108PD (REBP1971), S1274108 (REBQ1883), S3274355 (REBQ1884), S3274335 (REBQ1885), S3274108D (REBQ1886), S1274108D (REBQ1887), S9274108PD (REBQ1888), S0274108PD (REBQ1889), S3274335P (REBQ1890), S1274118 (REBQ1891), S9274108D (REBQ1892), S1274108D (REBR0160), S1274108D (REBN1883), S1173108D (REBN2329), S9173108D (REBN2332), S1274108PD (REBN2378), S9274108PD (REBN2379), S1173108PD (REBP0948), S3173335 (REBP0949), S3173355 (REBP0950), S9173108PD (REBP1020), s3173355p (REBP1023), S3173335P (REBP1024), S3274108D (REBQ1276), S1173108 (REBQ1882), S1173108 (REBQ2473), S1173108D (REBQ2474), S9173108D (REBQ2475), S1173108PD (REBQ2479), S0173108PD (REBR1399), S1173108PD (REBS0167), S3274108D (REBQ2476), S3173108PD (REBQ2480), S3173355 (REBQ2586), S3173335 (REBQ2587), s3173355p (REBQ2590), s3173108d (REBQ2592), S1173108PD (REBR1398), S0173108PD (REBS2081), S1274108D , REBS0062), S1274108D (REBS1035), S1274108D (REBS1567), S1274108D (REBR1347), S9274108D (REBQ2477), S3274355P (REBS0026), S3274335P (REBS0534), S3274335 (REBS0574), S3274355P (REBS1557), S3274355P (REBS2011), S3274355P (REBS2550), S1274108D (REBS2562), S1274118 (REBT0078), s1274108pd (REBT0146), S1274108PD (REBT0146), S3274355P (REBT0168), S3274355P (REBT0204), CK000765 (REBT0255), S1274108 (REBT0682), S0274108PD (REBT0697), S9274108D (REBT0754), S1274108D (REBT0755), S1274108PD (REBT0757), S3274335 (REBT0805), CK000765 (REBT0843), S9274108PD (REBT1446), S9274108 (REBT1885), CK000767 (REBT1889), S3274108PD (REBT2364), S9274108D (REBU0059), CK000767 (REBW1216), S3274355 (REBR1259), S3173355 (REBR1261), S1274108PD (REBR1345), S3274108PD (REBR1401), S1274108D (REBS0781), S3274355P (REBS0843), S1274108D (REBS1033), S1274108D (REBS2070), CK000759 (REBS2133), S3173355P (REBT0101), S3173335P (REBT0116), S3173355P (REBT0737), S3274335 (REBT1585), S1274108D (REBT1602), S3274355 (REBT1802), S3274108 (REBT1884), ck000770 (REBU0664), S9274108D (REBU0691), S1173108 (REBU1584), S9274108D (REBU1639), S1173108D (REBV0536), S1274108D (REBW0268), S3173108D (REBR1014), S1173108D (REBS1988), S3173108D (REBT0147), S9173108PD (REBT0234), S1173108D (REBT0240), S1173108D (REBT1398), S3173355P (REBT1801), S3173108 (REBT1882), S9173108 (REBT1883), S3173355P (REBT1948), S3173355P (REBU0082), S1173108D (REBU0293), S1173108D (REBU0690), S3173355P (REBU0714), S0173108PD (REBU1224), S3173355P (REBU1265), s9173108pd (REBU2255), S3173335P (REBV0052), S3173118 (REBV1102), S9173118 (REBV1103), S1173108PD (REBW1843), S3173108PD (REBX1664), s1173108 (REBT0688), S1274108D (REBU0296), S1274108D (REBU0765), S1274108 (REBU1047), S1274108D (REBU1271), S9173108D (REBU1640), S1274108PD (REBW0728), s3173355 (REBW1532), S1173108 (REBX1560), S1173108PD (REBX1663), S3173118 (REBY0008), S1173108D (REBY0782), S1173108PD (REBY0832), S0173108PD (REBY1316), S1173108 (REBY1472), S1173108D (REBY2516), CK000283A (REBV1203), S3173108PD (REBS0171), S3274108PD (REBT0233), S9274108PD (REBT1340), S3274108PD (REBT1403), S1274108 (REBT1846), s1274108 (REBT2073), S9274118 (REBU0001), S9274118 (REBU0002), S3274355 (REBU0083), S3274108D (REBU0297), S9274108PD (REBU0694), S3274355 (REBU0734), S3274355 (REBV0533), S3274118 (REBV1104), CK000114A (REBV2151), S1274108PD (REBV2327), S1274108D (REBW1840), S1274108D (REBX0325), CK000770 (REBX0460), S1274108D (REBX1581), CK000759 (REBX2038), S1274108D (REBX2436), S1274108D (REBX0326), S1274108PD (REBY0780), S1274108D (REBY0835), S1274108 (REBY1004), S1274108D (REBZ0814), CK000114A (REBZ1268), S1274108D (REBV1690), CK000770 (REBV2233), S1274108D (REBV2326), S1274108D (REBW0782), CK000759 (REBW1228), S1274108D (REBW1841), S1274108D (REBW1988), CK000767 (REBX0441), CK000759 (REBX0457), S1274108PD (REBX1122), S1274108D (REBX1240), S1274108PD (REBY1239), S1173108D (REBZ0958), S1173108PD (REBZ1627), S1274108D (RECN0085), S1274108PD (RECN0939), S1274108 (RECN1068), CK000114A (RECQ0041), S3173335 (REBY1941), S1173108D (REBY1982), S9173108PD (REBZ0816), S1173108D (REBZ0838), S1173108PD (REBZ0840), S3173108PD (REBZ0960), S1173108PD (RECN0088), S1173108D , RECN1199), s1173108pd (RECN1200), S3173108PD (RECN1406), S1173108D (RECN2175), S1173108PD (RECN2177), S0173108PD (RECN2678), S1173108D (RECP0357), S1173108PD (RECP0594), S1173108 (RECP1895), S3173355 (RECP2311), S1274108D (REBY0518), S3274108D (REBY0834), S1274108PD (REBY2518), S1173108PD (REBY2740), S9274108PD (REBZ0815), s1274108pd (REBZ1718), S1274108PD , RECN0027), S1274108PD (RECN2563), S3274108D (REBX2246), S3274335 (REBY0879), S1274108D (REBY2013), S3274335 (RECN0888), S1274108D (RECN0938), S1274108D (RECN2573), S3274108D (RECN2679), S9274108D (RECP0358), S1274108D (RECP0590), S3274355 (RECP0846), S1274108PD (RECP2482), S3274108D (RECQ0417), S3274108D (REBY2517), S1274108D (REBY2739), S3274355 (REBZ0494), S3274335 (REBZ0681), S3274108D (REBZ0839), S1274108D (REBZ0885), S1274108D (REBZ1625), S1274108 (REBZ1678), S1274108D (RECN1381), S1274108D (RECN2461), S3274108D (RECP0591), S9274108PD (RECP1511), S3274355 (REBZ1826), S3274108D (RECN0023), S9274108D (RECN0084), S9274108D (RECN1403), S1173108 (REBY1170)
- 510(k) numbers
- ["K162443"]
- PMA numbers
- not on file
- Product code
- LJS
- Firm FEI
- 3006260740
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28