Bard Access Systems Inc.: PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.

Recall number
Z-0536-2018
Recalling firm
Bard Access Systems Inc.
Product
PowerPICC Provena Catheters, as 3 Fr. Single lumen and 4 Fr. Dual Lumen. Packaged in a poly tray with Tyvek lid within a header bag configuration or a breather bag configuration.
Status
Terminated
Initiated
2017-12-28
Posted
2018-02-06
Terminated
2021-06-14
Reason
Possible leaks on the 3 Fr. Single lumen and 4 Fr. Dual lumen PowerPICC catheters at the luer hub extension leg junction.
Root cause
Labeling design
Action
Customers were notified via letter on approximately 12/28/2017. Instructions include a reminder that if additional extension sets are attached to the luers of the catheters, they should be appropriately secured as per hospital recommended securement protocol, to refrain from having unsupported extension sets that could put stress on the luer extension leg junction of the catheter, and to notify customers if the product was further distributed. Also, the notification contained a statement that Bard Access is in the process of updating their Instructions for Use to include a statement regarding the importance of properly securing additional extension sets per hospital recommended securement protocol. On 04/11/2018, the recall was expanded and customers were notified via letter, which explained that the previous recall has been expanded to involve return of affected product as well as include product distributed from February 2017 through March 2018. Instructions included to examine inventory for affected product, quarantine and immediately discontinue distribution of the product, complete and return the Business Recall Response Form, arrange for the return of affected product, and notify customers if the product has been further distributed. For distributors, Bard Access Systems requested a customer list so that they could notify their customers.
Quantity
53,699 units total
Distribution
Distribution US nationwide.
Lot, serial or model codes
Product Code (Lot/Serial No.): S1274108D (REBN2334), S9274108D (REBN2336), S3274108PD (REBP1471), S3274335 (REBP1580), S3274355 (REBP1581), S3274355P (REBP1582), S3274335P (REBP1583), S1274108PD (REBP1971), S1274108 (REBQ1883), S3274355 (REBQ1884), S3274335 (REBQ1885), S3274108D (REBQ1886), S1274108D (REBQ1887), S9274108PD (REBQ1888), S0274108PD (REBQ1889), S3274335P (REBQ1890), S1274118 (REBQ1891), S9274108D (REBQ1892), S1274108D (REBR0160), S1274108D (REBN1883), S1173108D (REBN2329), S9173108D (REBN2332), S1274108PD (REBN2378), S9274108PD (REBN2379), S1173108PD (REBP0948), S3173335 (REBP0949), S3173355 (REBP0950), S9173108PD (REBP1020), s3173355p (REBP1023), S3173335P (REBP1024), S3274108D (REBQ1276), S1173108 (REBQ1882), S1173108 (REBQ2473), S1173108D (REBQ2474), S9173108D (REBQ2475), S1173108PD (REBQ2479), S0173108PD (REBR1399), S1173108PD (REBS0167), S3274108D (REBQ2476), S3173108PD (REBQ2480), S3173355 (REBQ2586), S3173335 (REBQ2587), s3173355p (REBQ2590), s3173108d (REBQ2592), S1173108PD (REBR1398), S0173108PD (REBS2081), S1274108D , REBS0062), S1274108D (REBS1035), S1274108D (REBS1567), S1274108D (REBR1347), S9274108D (REBQ2477), S3274355P (REBS0026), S3274335P (REBS0534), S3274335 (REBS0574), S3274355P (REBS1557), S3274355P (REBS2011), S3274355P (REBS2550), S1274108D (REBS2562), S1274118 (REBT0078), s1274108pd (REBT0146), S1274108PD (REBT0146), S3274355P (REBT0168), S3274355P (REBT0204), CK000765 (REBT0255), S1274108 (REBT0682), S0274108PD (REBT0697), S9274108D (REBT0754), S1274108D (REBT0755), S1274108PD (REBT0757), S3274335 (REBT0805), CK000765 (REBT0843), S9274108PD (REBT1446), S9274108 (REBT1885), CK000767 (REBT1889), S3274108PD (REBT2364), S9274108D (REBU0059), CK000767 (REBW1216), S3274355 (REBR1259), S3173355 (REBR1261), S1274108PD (REBR1345), S3274108PD (REBR1401), S1274108D (REBS0781), S3274355P (REBS0843), S1274108D (REBS1033), S1274108D (REBS2070), CK000759 (REBS2133), S3173355P (REBT0101), S3173335P (REBT0116), S3173355P (REBT0737), S3274335 (REBT1585), S1274108D (REBT1602), S3274355 (REBT1802), S3274108 (REBT1884), ck000770 (REBU0664), S9274108D (REBU0691), S1173108 (REBU1584), S9274108D (REBU1639), S1173108D (REBV0536), S1274108D (REBW0268), S3173108D (REBR1014), S1173108D (REBS1988), S3173108D (REBT0147), S9173108PD (REBT0234), S1173108D (REBT0240), S1173108D (REBT1398), S3173355P (REBT1801), S3173108 (REBT1882), S9173108 (REBT1883), S3173355P (REBT1948), S3173355P (REBU0082), S1173108D (REBU0293), S1173108D (REBU0690), S3173355P (REBU0714), S0173108PD (REBU1224), S3173355P (REBU1265), s9173108pd (REBU2255), S3173335P (REBV0052), S3173118 (REBV1102), S9173118 (REBV1103), S1173108PD (REBW1843), S3173108PD (REBX1664), s1173108 (REBT0688), S1274108D (REBU0296), S1274108D (REBU0765), S1274108 (REBU1047), S1274108D (REBU1271), S9173108D (REBU1640), S1274108PD (REBW0728), s3173355 (REBW1532), S1173108 (REBX1560), S1173108PD (REBX1663), S3173118 (REBY0008), S1173108D (REBY0782), S1173108PD (REBY0832), S0173108PD (REBY1316), S1173108 (REBY1472), S1173108D (REBY2516), CK000283A (REBV1203), S3173108PD (REBS0171), S3274108PD (REBT0233), S9274108PD (REBT1340), S3274108PD (REBT1403), S1274108 (REBT1846), s1274108 (REBT2073), S9274118 (REBU0001), S9274118 (REBU0002), S3274355 (REBU0083), S3274108D (REBU0297), S9274108PD (REBU0694), S3274355 (REBU0734), S3274355 (REBV0533), S3274118 (REBV1104), CK000114A (REBV2151), S1274108PD (REBV2327), S1274108D (REBW1840), S1274108D (REBX0325), CK000770 (REBX0460), S1274108D (REBX1581), CK000759 (REBX2038), S1274108D (REBX2436), S1274108D (REBX0326), S1274108PD (REBY0780), S1274108D (REBY0835), S1274108 (REBY1004), S1274108D (REBZ0814), CK000114A (REBZ1268), S1274108D (REBV1690), CK000770 (REBV2233), S1274108D (REBV2326), S1274108D (REBW0782), CK000759 (REBW1228), S1274108D (REBW1841), S1274108D (REBW1988), CK000767 (REBX0441), CK000759 (REBX0457), S1274108PD (REBX1122), S1274108D (REBX1240), S1274108PD (REBY1239), S1173108D (REBZ0958), S1173108PD (REBZ1627), S1274108D (RECN0085), S1274108PD (RECN0939), S1274108 (RECN1068), CK000114A (RECQ0041), S3173335 (REBY1941), S1173108D (REBY1982), S9173108PD (REBZ0816), S1173108D (REBZ0838), S1173108PD (REBZ0840), S3173108PD (REBZ0960), S1173108PD (RECN0088), S1173108D , RECN1199), s1173108pd (RECN1200), S3173108PD (RECN1406), S1173108D (RECN2175), S1173108PD (RECN2177), S0173108PD (RECN2678), S1173108D (RECP0357), S1173108PD (RECP0594), S1173108 (RECP1895), S3173355 (RECP2311), S1274108D (REBY0518), S3274108D (REBY0834), S1274108PD (REBY2518), S1173108PD (REBY2740), S9274108PD (REBZ0815), s1274108pd (REBZ1718), S1274108PD , RECN0027), S1274108PD (RECN2563), S3274108D (REBX2246), S3274335 (REBY0879), S1274108D (REBY2013), S3274335 (RECN0888), S1274108D (RECN0938), S1274108D (RECN2573), S3274108D (RECN2679), S9274108D (RECP0358), S1274108D (RECP0590), S3274355 (RECP0846), S1274108PD (RECP2482), S3274108D (RECQ0417), S3274108D (REBY2517), S1274108D (REBY2739), S3274355 (REBZ0494), S3274335 (REBZ0681), S3274108D (REBZ0839), S1274108D (REBZ0885), S1274108D (REBZ1625), S1274108 (REBZ1678), S1274108D (RECN1381), S1274108D (RECN2461), S3274108D (RECP0591), S9274108PD (RECP1511), S3274355 (REBZ1826), S3274108D (RECN0023), S9274108D (RECN0084), S9274108D (RECN1403), S1173108 (REBY1170)
510(k) numbers
["K162443"]
PMA numbers
not on file
Product code
LJS
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register