CareFusion 303, Inc.: Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.

Recall number
Z-0534-2014
Recalling firm
CareFusion 303, Inc.
Product
Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.
Status
Terminated
Initiated
2013-12-05
Posted
2013-12-20
Terminated
2016-06-02
Reason
CareFusion is recalling the Alaris PC unit (model 8000 and 8015) because the Alaris PC unit will produce an audible alarm and the Alaris PC unit will display a System Error message.
Root cause
Software design
Action
CareFusion sent an Urgent Medical Device Recall Notification letter dated December 5, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to contact CareFusion Customer Advocacy at (888) 812-3266, if they experienced a SYSTEM ERROR on the Alaris PC unit. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Customers are instructed to complete and return the enclosed Customer Response Card to expedite the corrective action process. Customers with questions were instructed to contact CareFusion Support Center at 888-562-6018, or for Technical Support call 888-812-3229.
Quantity
4824 total units (4242 units in US)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, and Australia.
Lot, serial or model codes
not available
510(k) numbers
["K041241","K051641"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2016493
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register