Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: BIOMOD PORCOAT HA GLENTRAY; Custom Shoulder Products Item Nos. PM0000097 PM555140 PM555147 PM555319 PM555418 113867 113865 113863 Product Usage: Total Shoulder Replacement

Recall number
Z-0533-2020
Recalling firm
Zimmer Biomet, Inc.
Product
BIOMOD PORCOAT HA GLENTRAY; Custom Shoulder Products Item Nos. PM0000097 PM555140 PM555147 PM555319 PM555418 113867 113865 113863 Product Usage: Total Shoulder Replacement
Status
Terminated
Initiated
2019-08-21
Posted
not on file
Terminated
2021-07-21
Reason
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
Root cause
Environmental control
Action
On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
Quantity
214741
Distribution
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Lot, serial or model codes
LOTS  584440 823130 966710 653720 191230 361690 917170 993450 009590 971720 197730 421300 970590 799740 713840 265680 924350 729030 843020 729020 034400 279060 958550 451520 040360 364420 399470 417590 446550 449550 489790 572090 587200 609410 616720 965280 975560 996130 873200 287000 951480 031170 659320 830600 297280 407350 907020 751730 641070 713850 729050 729040 960150 957220 451540 299920 564400 150450 770600 364540 146250 761320 957590 868620 537430 810590 364560    UDI  (01)00887868226441(17)270911(10)823130 (01)00880304804180(17)251120(10)966710 (01)00880304860001(17)270228(10)191230 (01)00880304443044(17)230220(10)265680 (01)00880304443044(17)230315(10)729030 (01)00880304443044(17)230319(10)843020 (01)00880304443044(17)230328(10)729020 (01)00880304443044(17)230430(10)034400 (01)00880304443044(17)230520(10)279060 (01)00880304443044(17)230816(10)958550 (01)00880304443044(17)240212(10)451520 (01)00880304434127(17)230319(10)729050 (01)00880304434127(17)230328(10)729040 (01)00880304434127(17)230408(10)960150 (01)00880304434127(17)230816(10)957220 (01)00880304434127(17)240221(10)451540 (01)00880304434127(17)240714(10)299920 (01)00880304434127(17)250319(10)564400 (01)00880304434127(17)250331(10)150450 (01)00880304434127(17)260118(10)770600 (01)00880304433496(17)230816(10)957590 (01)00880304433496(17)240327(10)868620 (01)00880304433496(17)240521(10)537430 (01)00880304433496(17)240612(10)810590
510(k) numbers
["K043100"]
PMA numbers
not on file
Product code
MBF
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register