Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp. d.b.a. Integra Pain Management: Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local or spinal anesthetics. The BD LOR syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.

Recall number
Z-0533-2014
Recalling firm
Integra LifeSciences Corp. d.b.a. Integra Pain Management
Product
Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local or spinal anesthetics. The BD LOR syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.
Status
Terminated
Initiated
2013-10-28
Posted
2013-12-17
Terminated
2014-01-29
Reason
Integra LifeSciences Corporation is recalling certain lots of Integra Kits for Epidural Anesthesia Administration as a result of a Product Advisory Notice from BD Medical for the LOR syringes indicating the plunger of the syringe may stall or stick when travelling within the syringe.
Root cause
Component design/selection
Action
BD Medical sent a Product Advisory Notice letter dated August 23, 2013 to Iaffected consignees by via traceable courier. The letter identified the affected product, problem and actions to be taken. Distributor have been asked to forward copies of the notification to their customers. For questions call BD Customer Service at 1-888-237-2762.
Quantity
359,250
Distribution
USA Nationwide Distribution and one foreign consignee in Mursalat.
Lot, serial or model codes
Kit Cat No: 10-2433,13-2595,13-2616,13-2628,3104195,3403061,3403150,3403316,3403579,3403727,3403776,3403813,3403989,3404131,3404269,3404293,3404417,3404468,3404578,3503559,CUS025,CUS1121-01,CUS1164-02,CUS1202-04,CUS1298-03,CUS1376,CUS1376,CUS1400-01,CUS1406-03,CUS1415-04,CUS1568-01,CUS1664-01,CUS1696,CUS1723,CUS1760,CUS1773-01,CUS1797,CUS279-02,CUS416,CUS597,CUS628-02,CUS755,PISLL7,RSM394,RSM395,SS1246,SS2036-01,
510(k) numbers
["K945484"]
PMA numbers
not on file
Product code
CAZ
Firm FEI
1000138491
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register