FDA Medical Device Recalls (openFDA)
MEDLINE INDUSTRIES, LP - Northfield: (1) Medline Open Heart CDS, REF CDS840132O, packaged 1 kit/case, non-sterile; and (2) Medline Dr. Davies CV, REF DYNJ903016G, packaged 1 kit/case, non-sterile.
- Recall number
- Z-0532-2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- (1) Medline Open Heart CDS, REF CDS840132O, packaged 1 kit/case, non-sterile; and (2) Medline Dr. Davies CV, REF DYNJ903016G, packaged 1 kit/case, non-sterile.
- Status
- Open, Classified
- Initiated
- 2024-10-15
- Posted
- 2024-11-25
- Terminated
- not on file
- Reason
- Some of the suture packages within the convenience kits contain the incorrect needle type/size and suture diameter/length.
- Root cause
- Mixed-up of materials/components
- Action
- The recalling firm issued letters on and dated 10/15/2024 via email and first-class mail explaining the issue, potential risk, and informing the consignee stickers are to be provided upon request to be placed onto the Medline pack instructing that the affected component is to be removed and replaced with an item in inventory. Required actions were included which were to (1) Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately; (2) Use the link and information in the letter to complete the response form. List the quantity of affected product in inventory on the form. Upon receipt of the submitted response form, the consignee will receive over-labels to place on affected inventory with instructions for the consignee staff to remove the affected component prior to using the kit. The labels will be provided via FedEx overnight; (4) If the consignee is a distributor or has resold or transferred this product to another company or individual, the consignee is to notify them of the recall communication.
- Quantity
- 408 kits
- Distribution
- US Nationwide distribution in the states of AL, GA, MA, NJ, and PA.
- Lot, serial or model codes
- (1) REF CDS840132O - Lot number 24BBF701, exp. 10/31/2024, UDI-DI each 10195327245276, UDI-DI case 40195327245277; (2) REF DYNJ903016G - Lot numbers 24DBP801, exp. 8/31/2025; and 24CBJ749, exp. 11/30/2024, UDI-DI each - 10195327657154, UDI-DI case - 40195327657155.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OEZ
- Firm FEI
- 1417592
- Firm city
- Northfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28