FDA Medical Device Recalls (openFDA)

Trans American Medical, Inc.: Spectre Wireless Footswitch/Handswitch system Model 98/99, Model FDS, Model FDB. The Spectre Wireless Encrypted Footswitch/Hand Switch System is a prescription use system and intended only for use with the system indicated in the model number

Recall number
Z-0532-2014
Recalling firm
Trans American Medical, Inc.
Product
Spectre Wireless Footswitch/Handswitch system Model 98/99, Model FDS, Model FDB. The Spectre Wireless Encrypted Footswitch/Hand Switch System is a prescription use system and intended only for use with the system indicated in the model number
Status
Terminated
Initiated
2013-08-22
Posted
2014-01-03
Terminated
2015-12-29
Reason
TransAmerican Medical Imaging is recalling certain lots of TMI Spectre Wireless Footswitch/Hand switch for GE OEC 9800/9900 and Philips FDS and FDB Cath Labs due to potential safety issue related to the use of this product producing uncommanded X-ray.
Root cause
Component design/selection
Action
TransAmerican Medical sent an Urgent Medical Device Correction letter dated August 26, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that all potential users in their facility are made aware of the safety notification and the recommended actions. If the customers system does produce an uncommanded X-ray, turn the system off immediately. Remove the Spectre Wireless Footswitch and receiver and reinstall the standard OEM footswitch. Follow proper shutdown/system reboot instructions as indicated in the system operator's manual. Customers will be contacted to make arrangements for replacement of their TMi Spectre Wireless Footswitch at no charge. Customers will be provided a new TMI Spectre Wireless Footswitch and instructions on how to return the affected footswitch. Customers with questions were instructed to call 801-796-7335, ext 203.
Quantity
68
Distribution
US Distribution: including states of: WA and UT.
Lot, serial or model codes
Serial numbers NS 10324 through NS 10392
510(k) numbers
["K092713"]
PMA numbers
not on file
Product code
IZL
Firm FEI
1000219700
Firm city
Lindon
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register