Varian Medical Systems, Inc. Oncology Systems: Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, Reference/FSCA Identifier: CP-09798; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA

Recall number
Z-0531-2013
Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Product
Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Affected Versions: ARIA version 11.0 below 11.0.55, Build 11.0.28 and 11.0.34, Reference/FSCA Identifier: CP-09798; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
Status
Terminated
Initiated
2012-11-05
Posted
2012-12-17
Terminated
2014-03-03
Reason
Varian Medical Systems has identified an anomaly with the ARIA Oncology Information System Radiation Oncology [ARIA RO] DICOM import/export functionality where a 360 degree arc field may convert to static field during DICOM export or import.
Root cause
Software design
Action
Varian sent an Urgent Medical Device Correction Urgent Field Safety Notice dated November 6, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that the plan was not converted unintentionally during import/export. Customers would be contacted by a Customer Service Representative when the correction was available to schedule installation on their system. Customers with questions were instructed to contact their local Varian Medical Systems Customer Support District or Regional Manager. US and Canada 1-888-827-4265. For questions regarding this recall call 650-424-5731.
Quantity
264 units
Distribution
Worldwide Distribution
Lot, serial or model codes
CODES:  HITOO28, HIT0963, HITl780, HIT242S, HIT4118RI, HIT0632M, HITIS87R4, HIT2208RS, HID088R6, HITOO32, HIT0983M, HITl800, HIT2434, HIT4263, HlT0632Rl, HlTlS89M, HIT2212, HIT3097, HITOO49M, HIT0983RI, HITI8S4, HIT244I, HIT438I, HIT0632R2, HITlS89Rl, HlT2237M, HIT316S, HITOl16M, HIT0983R2, HITl874, HIT2477M, HIT4439RI, HIT0668, HITIS89R2, HlT2237RI, HIT3206, HITOI16RI, HIT0998, HITl898, HIT2477RI, HIT4SSS, HIT0691M, HITIS89R3, HIT2267M, HIT3267, HIT0134, HITI036, HITl9S8, HlT2477R2, HIT4S92, HIT0691RI, HlTlS89R4, HlT2267RI, HIT3268, HITOISS, HITl119M, HITl968, HIT2477R3, HIT4627, HIT0691R2, HITlS89RS, HIT227SM, HIT3308, HITOI70M, HITII19RI, HIT2037M, HIT2SI2M, HIT477I, HIT0691R3, HITlS89R6, HIT227SRI, HIT3438M, HITOI70RI, HITI139, HIT2037RI, HIT2SI2RI, HlT4800, HIT0764, HITlS89R7, HIT227SR2, HIT3S13, HITOI70R2, HITll49M, HIT2037RII, HIT2S22, HIT4884, HIT0816, HITIS89R8, HlT227SR3, HIT3679, HITOI70R3, HITll49Rl, HIT2037R12, HlT2S46, HIT4888M, HIT0828, HITIS89R9, HIT227SR4, HIT3764RI, HlTOl82, HITI184Al, HIT2037RI3, HIT2S63M, HIT4888RI, HIT08S2, HITI622, HlT2291M, HlT377S, HITOl83, HlTll84M, HIT2037R14, HIT2S63RI, HIT4888R2, HIT090lM, HITI660M, HIT229IRI, HlT3796, HITOl91M, HITll84RI, HIT2037RIS, HIT2S63R2, HIT4888R3, HIT090lRI, HlTl660R2, HIT2291R2, HID910, HITOl91RI, HITI229, HIT2037R16, HIT2S63R3, HIT4894, HIT0926, HITl660R3, HlT2291R3, HIT3927M, HITOI91R2, HITI277, HIT2037R17, HIT2S63R4, HIT4896, HIT0933M, HlTI671, HIT2291R4, HlT3927RI, HIT0227, HITl363M, HlT2037R2, HIT2S91, HIT4903, HIT0933RI, HITl693, HIT229IRS, HlT3927R2, HIT0337M, HITl363RI, HlT2037R3, HIT26IS, HlT4906, HIT0933R2, HITl726, HIT2291R6, HIT394S, HIT0337RI, HITl4I2, HlT2037R4, HIT2643, HIT4909, HIT0933R3, HITl7S7, HlT2293, HlT3967, HIT0337R2, HITl430M, HIT2037RS, HlT264S, HlT4932M, HIT0933R4, HITl764M, HIT2336M, HIT4014, HIT0337R3, HITl430RI, HIT2037R6, HIT2834, HIT4932RI, HIT0933RS, HITI764RI, HIT2336RI, HIT4107, HIT0339M, HITI430R2, HIT2037R7, HIT286S, HIT4938, HIT0933R6, HITl767M, HIT2346M, HIT41l6, HIT0339RI, HITl430R3, HIT2037R8, HIT2894M, HIT4942M, HIT0933R7, HITI767RI, HIT2346RI, HIT41I8M, HIT0341M, HITI470M, HIT2037R9, HIT2894Rl, HIT4942RI, HIT0341RI, HITl470RI, HIT2067, HIT2894R2, HIT4947, HIT0342, HITl470R2, HIT209S, HIT2894R3, HIT4974, HIT0383, HITI498, HIT2102, HlT2943, HIT4975, HIT0397, HITlSI4, HIT2140, HlT29S3, HIT4982, HIT0432M, HITlS23M, HIT2I80, HIT2997, HITSOII, HIT0432RI, HITlS23RI, HIT2188, HIT3088M, HITS048, HIT0432R2, HITIS7S, HIT2208M, HIT3088RI, HITS061, HIT0468M, HITlS87M, HIT2208Rl, HIT3088R2, HITS088, HIT0468RI, HlTlS87RI, HIT2208R2, HIT3088R3, HITS089, HITOS3S, HlTlS87R2, HIT2208R3, HIT3088R4, HITSI16, HITOSSI, HITIS87R3, HIT2208R4, HIT3088R5, HITSl41
510(k) numbers
["K093527"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register