Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo ultrasonic air sensor, 1/4" by 1/16" (6.4 mm x 1.6 mm), Black, for use on Terumo perfusion systems 1, 8000 and 9000; Catalog No: 5785, Terumo Cardiovalvular Systems, Ann Arbor, MI 48103

Recall number
Z-0531-2008
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo ultrasonic air sensor, 1/4" by 1/16" (6.4 mm x 1.6 mm), Black, for use on Terumo perfusion systems 1, 8000 and 9000; Catalog No: 5785, Terumo Cardiovalvular Systems, Ann Arbor, MI 48103
Status
Terminated
Initiated
2007-09-19
Posted
2008-01-24
Terminated
2009-07-07
Reason
False Alarm: The air sensor may malfunction and trigger false alarms, and may continue to alarm thus preventing the device from being reset.
Root cause
Process control
Action
Consignees were notified of the defect/issue and what to do if it were to occur (short term work around) via an Urgent Medical Device Removal letter dated 9/14/07. As the permanent fix users were informed the sensor will be replaced when replacements become available.
Quantity
232
Distribution
Worldwide: USA, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Costa Rica, Dominican Republic, Equador, Germany, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, Uruguay, Venezuela and Vietnam.
Lot, serial or model codes
Serial Numbers: 1106 through 1370 and 1376.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
KRL
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register