Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: AMS Artificial Urinary Sphincter Control Pump

Recall number
Z-0530-2025
Recalling firm
Boston Scientific Corporation
Product
AMS Artificial Urinary Sphincter Control Pump
Status
Open, Classified
Initiated
2024-10-28
Posted
2024-11-25
Terminated
not on file
Reason
Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
Root cause
Error in labeling
Action
Boston Scientific notified their consignees on about 10/28/2024 via letter. Consignees were instructed to immediately discontinue use and segregate the affected products, notify all personnel that need to be aware, notify customers if further distributed, arrange for the return of affected product and complete and return the Reply Verification Tracking Form (RVtF).
Quantity
16 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, KS, LA, MA, MI, MN, NY, OR, PA, TX, VA, WA and the countries of Argentina, Australia, Austria, Belgium, Chile, China, Costa Rica, Czech Republic, Finland, France, Germany, Ireland, Netherlands, Poland, Puerto Rico, South Korea, Spain, Sweden, Switzerland, and Taiwan.
Lot, serial or model codes
UPN Number: 72400098. GTIN Number: 00878953000688. Lot Number: 32487464. Serial Numbers: 1100366307, 1100366316, 1100366309, 1100366317, 1100366310, 1100366319, 1100366311, 1100366320, 1100366312, 1100366321, 1100366313, 1100366323, 1100366314, 1100366324, 1100366315, 1100366325. Expiration Date: 09/24/2028
510(k) numbers
not on file
PMA numbers
["P000053"]
Product code
EZY
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register