Skytron, Div. The KMW Group, Inc: Skytron Infinity Series surgical light with dual lightheads, 5 bulbs in each of 2 - 22' diameter lightheads, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Models IF2222, IF2222-B, IF2222B/PT, IF2222EL and IN2222EL.

Recall number
Z-0530-2007
Recalling firm
Skytron, Div. The KMW Group, Inc
Product
Skytron Infinity Series surgical light with dual lightheads, 5 bulbs in each of 2 - 22' diameter lightheads, Manufactured by Dai-ichi Shomei Co., Ltd., Made in Japan; Models IF2222, IF2222-B, IF2222B/PT, IF2222EL and IN2222EL.
Status
Terminated
Initiated
2007-01-08
Posted
2007-02-22
Terminated
2009-11-18
Reason
If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.
Root cause
Other
Action
The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Quantity
5,000 for all models.
Distribution
Nationwide and Canada.
Lot, serial or model codes
Each lighthead bears a unique serial number. For units with model numbers beginning with IF22, all serial numbers with the following numbers in the third and fourth position of the numerical series: 95, 96, 97, 98, 99 or 00 (examples: xx97xxxx or xx00xxxB). For units with model numbers beginning with IN22 , all serial numbers with the following numbers in the third and fourth positions of the numerical series: 92, 93 or 94 (example: xx93xxxx).
510(k) numbers
["K915285"]
PMA numbers
not on file
Product code
FTD
Firm FEI
1825014
Firm city
Grand Rapids
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register