FDA Medical Device Recalls (openFDA)
EOS Imaging: EOS System X- ray beam Digital radiography system used in general radiographic examinations.
- Recall number
- Z-0529-2014
- Recalling firm
- EOS Imaging
- Product
- EOS System X- ray beam Digital radiography system used in general radiographic examinations.
- Status
- Terminated
- Initiated
- 2013-11-07
- Posted
- 2014-01-09
- Terminated
- 2015-01-27
- Reason
- EOS imaging discovered during production internal testing that the X-ray beam collimation set up by the operator at the user interface of the acquisition workstation is different from the actual collimation of the X-ray beam operated by the EOS system during the X-ray acquisition.
- Root cause
- Device Design
- Action
- CDRH approves the following conditions for EOS Imaging SA: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator notes below what are to be included in the notification. The draft letter provided on November 28, 2013, is deficient for the following reasons: a. It does not provide a statement that all repairs will be made without charge (21 CFR 1003.21(3)). b. It does not provide instructions, with respect to the use of the product pending the correction of the defect (21 CFR 1003.21(1). Specifically, the letter should include directions for the user to determine if the collimation defect is currently present in their system, which would initiate immediate correction, rather than at the next semi-annual service. 2. The customer notification letter will be sent in the manner required by 21 CFR 1003.21(b). 3. The repairs outlined are to be completed by July 1, 2014. All systems under service contracts (which includes semi-annual service) would be visited between receipt of this letter and July 1, 2014. All repairs, including those to systems which are not under service contracts, are also expected to be completed by July 1, 2014. EOS may proceed with implementation of the CAP (subject to the conditions noted above. The customer notification letter may be sent out immediately upon addressing the above deficiencies. However, it is suggested that you provide a copy of the revised letter to FDA for approval prior to issuance. For further questions please call (678) 564-5400.
- Quantity
- 22 EOS systems are installed in US
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- EOS System
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MQB
- Firm FEI
- 3006165260
- Firm city
- Paris France
- Firm state
- not on file
- Firm country
- France
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28