Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Stryker Orthopaedics Sterile Knee Instrument, Patella Cutter; Stainless Steel, Disposable; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 33 mm size, One single unit to a package. Note: The 30 mm instruments were not distributed. The patella cutters are components of the Xcelerate Instruments and are used with additional devices to remove bone from the patella in preparation for a replacement implant. The Xcelerate Patella Reaming Instruments provide a means to prepare for implantation of either the Duracon or Scorpio Total Knee Systems patellar implants.

Recall number
Z-0529-2010
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Stryker Orthopaedics Sterile Knee Instrument, Patella Cutter; Stainless Steel, Disposable; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 33 mm size, One single unit to a package. Note: The 30 mm instruments were not distributed. The patella cutters are components of the Xcelerate Instruments and are used with additional devices to remove bone from the patella in preparation for a replacement implant. The Xcelerate Patella Reaming Instruments provide a means to prepare for implantation of either the Duracon or Scorpio Total Knee Systems patellar implants.
Status
Terminated
Initiated
2009-11-11
Posted
2009-12-15
Terminated
2010-08-30
Reason
Disposable Patella Cutters - 30 mm (catalog number 8000-0030, lot V31257001) were packaged and labeled as 33 mm Disposable Patella Cutters (catalog number 8000-0033, lot V31257001 and vice versa.
Root cause
Labeling mix-ups
Action
An "Urgent Product Recall" letter dated November 11, 2009 was sent to all affected Stryker branches, hospital risk management, hospital Chief or Orthopaedics and surgeons by Federal Express. The letter described the issue, potential hazards and risk mitigation factors and requested that the Recall Acknowledgement form be completed and returned. The Recall Acknowledgement form has information regarding the return of affected product to the firm. Direct questions to Stryker Orthopaedics by calling 1-201-831-5832.
Quantity
12
Distribution
Nationwide Distribution (MA, NJ, FL, TN, OR and NY).
Lot, serial or model codes
Catalog Number 8000-0033; Lot code V31257001.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTS
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register