Skip to the content

FDA Medical Device Recalls (openFDA)

Karl Storz Endoscopy America Inc: Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.

Recall number
Z-0527-2014
Recalling firm
Karl Storz Endoscopy America Inc
Product
Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.
Status
Terminated
Initiated
2013-11-13
Posted
2013-12-17
Terminated
2014-03-27
Reason
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
Root cause
Under Investigation by firm
Action
The firm, KARL STORZ, sent a "A SAFETY ALERT-ATTENTION NEEDED" letter dated November 8, 2013 to all their customers on November 13, 2013. The letter describes the product, problem and actions to be taken.The letter informs the customers that the endoscopes incorporated a lumen material that is only approved for use as an external surface material and not included in the STERRAD NX and 100NX FDA-cleared claims for use in flexible endoscope lumens. The customers were instructed to follow the recommended processing methods included in the letter. Customers with questions about the letter are instructed to contact the undersigned and the Technical Services at (800) 421-0837, ext 5350.
Quantity
141 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E
510(k) numbers
["K832010","K954797","K981458"]
PMA numbers
not on file
Product code
ECQ
Firm FEI
2020550
Firm city
Southbridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register