Skip to the content

FDA Medical Device Recalls (openFDA)

Hu-Friedy Mfg Co, Inc.: Dental Explorer #23 S/E, Mfr. Part Code/Product Code 100-3620, distributed under the Henry Schein label. Classification Name: 21CFR 872.4565; 510(k) exempt, Class I device, Device Listing # B057701. The product is labeled in part, HENRY SCHEIN 100-3620 Explorer-Single End #23 SE. +H658100362003. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. --- This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling).

Recall number
Z-0527-05
Recalling firm
Hu-Friedy Mfg Co, Inc.
Product
Dental Explorer #23 S/E, Mfr. Part Code/Product Code 100-3620, distributed under the Henry Schein label. Classification Name: 21CFR 872.4565; 510(k) exempt, Class I device, Device Listing # B057701. The product is labeled in part, HENRY SCHEIN 100-3620 Explorer-Single End #23 SE. +H658100362003. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. --- This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling).
Status
Terminated
Initiated
2005-01-07
Posted
2005-02-04
Terminated
2012-05-24
Reason
Potential for tip breakage at the terminal shank due to improper heat-treatment at the foreign plant. Several reports of tip breakage were received by the own label distributor, and forwarded to the importer of record & specification setter of the dental explorers.
Root cause
Other
Action
Hu-Friedy Mfg. Co., Inc. notified the Own Label Distributor of the recall on 1/7/05 by phone, and in writing on 1/10/05. Henry Schein, Inc. is having formal "URGENT: DEVICE RECALL" letters printed up which will be mailed to over 8100 of Schein's direct customers by First Class Mail on/about 1/21/05 to initiate the sub-recall of the dental explorers. Henry Schein will follow-up to non-responders by phone and/or fax.
Quantity
42,517 total units distributed (all date codes)
Distribution
Dental practitioners, hospitals, dental clinics and distributors located nationwide, in Puerto Rico and internationally.
Lot, serial or model codes
Date Codes/Lot Numbers: 0503, 0603, 0903, 1003, 1103, 1203. Barcode +H658100362003.
510(k) numbers
not on file
PMA numbers
not on file
Product code
EKB
Firm FEI
1416605
Firm city
Chicago
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register