NRT X-RAY A/S: 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)

Recall number
Z-0526-2022
Recalling firm
NRT X-RAY A/S
Product
1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)
Status
Open, Classified
Initiated
2022-01-21
Posted
not on file
Terminated
not on file
Reason
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
Root cause
Process control
Action
On 01/11/2022, NRT issued an Urgent Medical device correction, which includes the following action to be taken by the users. 1. Identified the affected accessories as described in the Urgent Medical Device Correction Letter. 2. Inspect the Footrest and Shoulder Support which use the same clamping mechanism and should be inspected in the same way. The clamping mechanism consists of two clamping profiles, one for each side of the tabletop. Both sides should be inspected. a. The friction rubber that must be inspected is placed inside the clamping profile. It is the thick rubber, which must be firmly in place. The rubber is glued on the profile, and you should look for the overall adhesion. If slightly pressed from each side and end, the rubber should not show signs of detachment from the profile. If the rubber is firmly in place, you may continue use the equipment. If the rubber is detaching along either side or end, do not use the accessory until the repair has been made. 3. Return the response card in page 4, with the results and date of the inspection and your signature, not later than February 15, 2022. to support@nrtxray.com. 4. Follow up will be made in regards to the repair in April 2022, which will be coordinated with US dealer Alpha Imaging in Willoughby, Ohio. For any question contact (300) 696-0004 or E-Mail at ksmith@alpha-imaging.com.
Quantity
72 units
Distribution
USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden
Lot, serial or model codes
1. Footrest, REF: 03282000, UDI: 05713464000336;  Serial Numbers: 903282000046, 903282000057, 903282000075 ,9032820000070, 903282000001, 903282000002, 903282000004 thru 903282000006, 903282000009, 903282000011, 903282000012 903282000015, 903282000016, 903282000018 thru 903282000020, 903282000023, 903282000024, 903282000025, 903282000026, 903282000028, 903282000032 thru 903282000034, 903282000036, 903282000038 thru 903282000042, 903282000044, 903282000045, 903282000047 thru 903282000051, 903282000053, 903282000055, 903282000056, 903282000058 thru 903282000061, 903282000063, 903282000064, 903282000067, 903282000068  2. Shoulder Support, REF: 03283000, UDI: 05713464000381 Serial Numbers: 903283000012, 903283000028, 903283000001 thru 903283000003, 903283000005, 903283000010, 903283000011, 903283000013 thru 903283000022, 903283000023, 903283000024, 903283000026
510(k) numbers
["K173631"]
PMA numbers
not on file
Product code
OWB
Firm FEI
1000188474
Firm city
Hasselager
Firm state
not on file
Firm country
Denmark

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register