Skip to the content

FDA Medical Device Recalls (openFDA)

Praxair Inc.: Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.

Recall number
Z-0525-2013
Recalling firm
Praxair Inc.
Product
Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.
Status
Terminated
Initiated
2012-11-16
Posted
2012-12-21
Terminated
2014-03-24
Reason
Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.
Root cause
Device Design
Action
The firm, PRAXAIR, sent a "MEDICAL DEVICE CORRECTION" letter dated November 26, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the Grab'n Go operating instructions provided in the letter which include: storing the Grab'n Go systems indoors in a safe area; never store a system in an elevated location; handle systems safely and avoid jarring or dropping the system and locate at the following web link: www.praxair.com/GNGinstructions. The customers were also instructed to ensure that all of the Grab'n Go units that were sent by Praxair are sent to Praxair for servicing and refilling after use, in accordance with their normal process. If you have any questions, contact your sales representative at 630-320-4452 or email: Thelma_Brantley@Praxair.com.
Quantity
320,451 units (255,690 US ; 55,884 Canada; 8877 OUS)
Distribution
Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.
Lot, serial or model codes
OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global)  Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed.
510(k) numbers
not on file
PMA numbers
not on file
Product code
CAN
Firm FEI
1220594
Firm city
Danbury
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register