Zimmer Inc.: Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1302-55M. Note: This lot was not distributed in the United States. The component is used in the stabilization of constructs in the BacFix and InCompass spinal fixation systems. It is used within a temporary construct that assists normal healing and is not intended to replace normal body structures.

Recall number
Z-0525-2010
Recalling firm
Zimmer Inc.
Product
Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1302-55M. Note: This lot was not distributed in the United States. The component is used in the stabilization of constructs in the BacFix and InCompass spinal fixation systems. It is used within a temporary construct that assists normal healing and is not intended to replace normal body structures.
Status
Terminated
Initiated
2009-09-21
Posted
2009-12-14
Terminated
2010-10-08
Reason
The titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal.
Root cause
Nonconforming Material/Component
Action
Consignees were notified by letter dated September 21, 2009 and instructed to return the affected product to Zimmer, Inc.
Quantity
13
Distribution
Worldwide Distribution -- Canada, Mexico and South Africa. (Product not distributed in the United States)
Lot, serial or model codes
Lot 55QJ.
510(k) numbers
["K002082"]
PMA numbers
not on file
Product code
KWP
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register