FDA Medical Device Recalls (openFDA)
Zyno Medical LLC: Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision physician or other certified healthcare professional.
- Recall number
- Z-0523-2025
- Recalling firm
- Zyno Medical LLC
- Product
- Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision physician or other certified healthcare professional.
- Status
- Open, Classified
- Initiated
- 2024-09-30
- Posted
- 2024-11-22
- Terminated
- not on file
- Reason
- For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- Root cause
- Software design (manufacturing process)
- Action
- On September 30, 2024 MEDICAL DEVICE NOTIFICATION letters were sent to customers. Mitigation: Zyno Medical is issuing this customer notification, advising customers that the expected User ID may be missing from completed therapies in the EMR Patient Record. Zyno Medical is planning to correct this issue through an update to the software. Recommended Actions: The user can locate the Patient Order in the Patient Record in the EMR and manually update the record to reflect the administrating clinician s User ID until the Z-800WF pump software version is updated. Follow the guidelines of your facility for performing manual updates in the EMR. Reporting Adverse Events: We strongly encourage healthcare providers and patients to promptly report any adverse events associated with the use of the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pump. Please send your reports to Zyno Medical via email to feedback@intuvie.com. Additionally, we urge you to report any negative side effects related to the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps to the FDA. You can submit your report online at www.fda.gov/medwatch or by calling 1-800-FDA-1088. Your cooperation in reporting these events is vital for continuous safety monitoring of medical devices. For information about the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps, including benefits and risks, refer to the Zyno Medical Z-800 Infusion Pump Instructions for Use (Z-800 800-IFU-6651, Z-800F 800-IFU-2602, Z-800W 800W-IFU-3614, and Z-800WF 800WF-IFU-4602) for complete description of the benefits and risks of using the Zyno Medical Z-800 family of pumps. Zyno Medical is committed to offering products which meet the highest quality standards and to continuous improvement in the interest of patient safety. We regret any inconvenience that this action may cause, but believe transparent communication will ensure you have timely, relevant information for managing your patients. If you
- Quantity
- 100 units
- Distribution
- Domestic: AL, FL, LA, MI, NE, NJ, OH, TX.
- Lot, serial or model codes
- Model/Catalog Number: Z-800WF; UDI-DI: 00814371020006; Serial Numbers: 500346, 500347, 500348, 500349, 500350, 500351, 500352, 500353, 500354, 500355, 500356, 500357, 500358, 500359, 500360, 500361, 500362, 500363, 500364, 500365, 500367, 500368, 500369, 500370, 500371, 500373, 500374, 500375, 500376, 500377, 500378, 500384, 500385, 500386, 500388, 500389, 500390, 500391, 500392, 500393, 500394, 502634, 503435, 503716, 503799, 503917, 503921, 504534, 504807, 504808, 504809, 504810, 504811, 504812, 504813, 504814, 504815, 504816, 504817, 504818, 504819, 504820, 504821, 504822, 504823, 504824, 504825, 504826, 504827, 504828, 504829, 504830, 504831, 504832, 504834, 504835, 504836, 504838, 505369, 505425, 505430, 505432, 505437, 505626, 505627, 505628, 505629, 505630, 505631, 505632, 505633, 505635, 505637, 505638, 505691, 505692, 505693, 506585, 506982, 506983.
- 510(k) numbers
- ["K130690","K100705"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 3006575795
- Firm city
- Natick
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28