Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Recall number
Z-0523-2015
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Status
Terminated
Initiated
2014-10-28
Posted
2014-12-06
Terminated
2015-07-20
Reason
It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
Root cause
Use error
Action
A safety advisory notice, dated October 28, 2014, was sent to users that identified the product, problem, and actions to be taken to prevent risk to patients and users. Siemens technicians werre instructed to contact customers to schedule a field modification.
Quantity
3
Distribution
US Nationwide Distribution in the state of: WV, NY, and FL..
Lot, serial or model codes
model number: 10140000
510(k) numbers
["K122286"]
PMA numbers
not on file
Product code
MUE
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register