FDA Medical Device Recalls (openFDA)
mobility unlimited inc: KayserBetten model IDA beds Product Usage: BED, AC-POWERED ADJUSTABLE HOSPITAL
- Recall number
- Z-0522-2014
- Recalling firm
- mobility unlimited inc
- Product
- KayserBetten model IDA beds Product Usage: BED, AC-POWERED ADJUSTABLE HOSPITAL
- Status
- Terminated
- Initiated
- 2013-04-11
- Posted
- 2013-12-17
- Terminated
- 2014-10-08
- Reason
- Some joints on the doors of a limited number of IDA beds may loosen or separate over time which can cause the doors to drop outward and cause injury.
- Root cause
- Device Design
- Action
- Mobility Unlimited sent a notification letters dated April 12, 2013 to affected customers The letter identified the affected product, problem and actions to be taken. The letter informed customers that a Retrofit Kits along with illustrated instructions will be sent. For questions call 1-800-574-7880.
- Quantity
- 33 beds
- Distribution
- Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, FL, GA, IL, MN, MO, NY, PA, TN, TX, VA, and Canada.
- Lot, serial or model codes
- Serial Nos.: 227.15.03s.-00.-07.10.0726/1998 228.15.03S.-00.-06.09.0562/1998 227.15.03s.-00.-07.10.0725/1998 227.15.03S.-55.-03.11.0141/1998 227.15.03S.-55.-03.11.0142/1998 226.15.03S.-00.-06.09.0557/1998 228.15.03S.-00.-06.09.0563/1998 225.02.03s.-35.-07.10.0733/1998 227.15.03S.-00.-06.09.0569/1998 227.15.03S.-00.-06.09.0572/1998 227.15.03S.-00.-06.09.0714/1998 228.15.03S.-00.-06.09.0559/1998 228.15.03S.-00.-06.09.0713/1998 227.15.03S.-00.-06.09.0567/1998 227.15.03S.-00.-06.09.0570/1998 228.15.03S.-00.-06.09.0560/1998 225.15.03s.-00.-07.10.0724/1998 228.15.03S.-00-.12.08.1366/1998 228.15.03S.-00.-06.09.0564/1998 225.15.03S.-00.-06.09.0565/1998 225.15.03S.-00.-06.09.0566/1998 225.02.03s.-35.-07.10.0685/1998 225.15.03s.-00.-07.10.0723/1998 227.15.03S.-00.-06.09.0571/1998 227.15.03S.-55.-03.11.0144/1998 225.02.03s.-35.-07.10.0686/1998 226.15.03S.-00.-06.09.0558/1998 225.15.03s.-00.-07.10.0732/1998 227.15.03S.-00.-12.08.1363/1998 227.15.03S.-55.-03.11.0143/1998 225.02.03s.-35.-07.10.0684/1998 225.15.03s.-00.-07.10.0731/1998 228.15.03S.-00.-06.09.0561/1998
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FNL
- Firm FEI
- 3006791363
- Firm city
- Allentown
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28