Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Leibinger USA: Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.

Recall number
Z-0521-2007
Recalling firm
Stryker Leibinger USA
Product
Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.
Status
Terminated
Initiated
2006-12-22
Posted
2007-02-27
Terminated
2008-06-04
Reason
Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.
Root cause
Other
Action
Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.
Quantity
26
Distribution
Nationwide.
Lot, serial or model codes
Lots G4C00F72HN, G5H00F72HN, G7T00F72HN, G8W00F72HN and G9700F72HN.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3005101424
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register