GE Medical Systems, LLC: Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR
- Recall number
- Z-0520-2025
- Recalling firm
- GE Medical Systems, LLC
- Product
- Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
- Status
- Open, Classified
- Initiated
- 2024-10-28
- Posted
- 2024-11-22
- Terminated
- not on file
- Reason
- Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Root cause
- Device Design
- Action
- GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to recall.60998@gehealthcare.com. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 2 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
- Lot, serial or model codes
- 1) UDI/DI To be provided, System ID 82427070262; 2) UDI/DI To be provided, System ID EX0133
- 510(k) numbers
- ["K163619","K213709","K142098"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 2183553
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28