Del Mar Reynolds Medical, Ltd.: Lifescreen Apnea is Lifescreen ECG Screening System, Version 3.31. Lifescreen Apnea is indicated for screening adults for the probability that they suffer from obstructive sleep apnea, mixed apnea or hypopnea, in order to evaluate the necessity for a polysomonographic examination. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.
- Recall number
- Z-0520-2012
- Recalling firm
- Del Mar Reynolds Medical, Ltd.
- Product
- Lifescreen Apnea is Lifescreen ECG Screening System, Version 3.31. Lifescreen Apnea is indicated for screening adults for the probability that they suffer from obstructive sleep apnea, mixed apnea or hypopnea, in order to evaluate the necessity for a polysomonographic examination. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep of at least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.
- Status
- Terminated
- Initiated
- 2011-11-11
- Posted
- 2012-01-11
- Terminated
- 2012-06-05
- Reason
- When Lifescreen is left open and a new recording is loaded from Sentinel Cardiology Information Management System, customers can get corruption in the data or a wrong ECG after the first 32 hours. Both recordings have to be longer than 32 hours for this to happen.
- Root cause
- Software design
- Action
- Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated November 11, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to close Lifescreen every time before loading another recording from the Sentinel database. Spacelabs is in the process of upgrading software and will contact customers to arrange for an update to affected systems. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
- Quantity
- total 523 units and 171 kits (34 units and 7 kits in US; 489 units and 164 kits outside US)
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, CO, CT, FL, GA, KY, LA, MA, MD, MI, MN, NC, NY, OH, TN, TX, UT, WV, and PUERTO RICO and the countries of AUSTRIA, BELGIUM, CHILE, CHINA, COLOMBIA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, ISRAEL, ITALY, KUWAIT, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, Slovakia, SPAIN, SWEDEN, SWITZERLAND, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM
- Lot, serial or model codes
- There is no Serial Numbers for 169 Kits. Serial Numbers for 519 units: 12038, 32288, 32300, 32302, 32327, 32329, 32330, 32389, 32410, 32411, 32415, 32417, 32418, 32470, 32471, 32472, 32473, 32474, 32475, 32476, 32478, 32479, 32480, 32481, 32482, 32483, 32484, 32485, 32486, 32487, 32488, 32532, 32546, 32547, 32548, 32549, 32550, 32551, 32552, 32553, 32554, 32557, 32558, 32561, 32567, 32568, 32569, 32570, 32648, 32649, 32650, 32651, 32652, 32653, 32655, 32656, 32657, 32658, 32659, 32676, 32677, 32678, 32679, 32680, 32681, 32682, 32683, 32684, 32686, 32687, 32712, 32713, 32714, 32715, 32716, 32717, 32718, 32719, 32720, 32721, 32722, 32723, 32726, 32742, 32759, 32790, 32791, 32797, 32798, 32799, 32800, 32801, 32802, 32803, 32804, 32805, 32806, 32814, 32821, 32822, 32823, 32824, 32860, 32861, 32862, 32863, 32864, 32865, 32866, 32867, 32868, 32869, 32870, 32871, 32872, 32873, 32877, 32878, 32880, 32881, 32944, 32945, 32946, 32947, 32948, 32949, 32950, 32951, 32952, 32954, 32955, 32956, 32957, 32958, 32959, 32960, 32961, 32963, 32977, 32979, 32980, 32982, 32983, 32984, 32986, 32987, 32989, 32990, 32991, 32992, 32993, 32994, 32995, 33015, 33017, 33058, 33095, 33096, 33097, 33098, 33099, 33100, 33101, 33102, 33103, 33104, 33105, 33106, 33107, 33108, 33113, 33114, 33115, 33116, 33117, 33118, 33119, 33120, 33121, 33122, 33123, 33124, 33132, 33135, 33158, 33159, 33160, 33161, 33162, 33164, 33165, 33166, 33167, 33184, 33185, 33186, 33188, 33189, 33190, 33200, 33201, 33213, 33219, 33221, 33240, 33241, 33242, 33243, 33244, 33245, 33254, 33255, 33257, 33258, 33259, 33260, 33261, 33262, 33263, 33264, 33265, 33267, 33268, 33269, 33270, 33322, 33323, 33324, 33326, 33327, 33328, 33329, 33330, 33336, 33337, 33338, 33339, 33343, 33347, 33348, 33352, 33355, 33386, 33387, 33388, 33391, 33392, 33393, 33394, 33396, 9310, 9520, 9885, 9933, 9934, IMP-002054, IMP-005010, IMP-005011, IMP-005013, IMP-005025, IMP-005027, IMP-005028, IMP-005029, IMP-005030, IMP-005031, IMP-005035, IMP-005036, IMP-005038, IMP-005039, IMP-005042, IMP-005043, IMP-005045, IMP-005046, IMP-005048, IMP-005049, IMP-005050, IMP-005062, IMP-005063, IMP-005065, IMP-005066, IMP-005068, IMP-005073, IMP-005077, IMP-005078, IMP-005087, IMP-005090, IMP-005091, IMP-005092, IMP-005094, IMP-005102, IMP-005108, IMP-005110, IMP-005111, IMP-005115, IMP-005141, IMP-005147, IMP-005151, IMP-005152, IMP-005160, IMP-005162, IMP-005165, IMP-005176, IMP-005218, IMP-005219, IMP-005237, IMP-005238, IMP-005253, IMP-005254, IMP-005258, LIFE-025878, LIFE-027365, LIFE-027527, LIFE-027528, LIFE-027866, LIFE-028314, LIFE-029149, LIFE-029150, LIFE-029718, LIFE-029719, LIFE-032784, LIFE-033181, LIFE-033791, PATH-000100, PATH-000101, PATH-000102, PATH-000103, PATH-000104, PATH-000105, PATH-000106, PATH-000107, PATH-006004, PATH-006005, PATH-006006, PATH-006007, PATH-006018, PATH-006019, PATH-006020, PATH-006039, PATH-006040, PATH-006041, PATH-006042, PATH-006043, PATH-006044, PATH-006045, PATH-006068, PATH-006069, PATH-006070, PATH-006071, PATH-006072, PATH-006073, PATH-006074, PATH-006075, PATH-006077, PATH-006099, PATH-006101, PATH-006102, PATH-006103, PATH-006104, PATH-006105, PATH-006106, PATH-006108, PATH-006109, PATH-006110, PATH-006112, PATH-006113, PATH-006114, PATH-006115, PATH-006120, PATH-006132, PATH-006158, PATH-006159, PATH-006160, PATH-006161, PATH-006162, PATH-006163, PATH-006164, PATH-006165, PATH-006166, PATH-006175, PATH-006182, PATH-006183, PATH-006184, PATH-006186, PATH-006188, PATH-006190, PATH-006237, PATH-006238, PATH-006239, PATH-006250, PATH-006251, PATH-006252, PATH-006253, PATH-006254, PATH-006255, PATH-006256, PATH-006258, PATH-006259, PATH-006299, PATH-006300, PATH-006301, PATH-006302, PATH-006306, PATH-006309, PATH-006314, PATH-006315, PATH-006316, PATH-006317, PATH-006318, PATH-006319, PATH-006320, PATH-006321, PATH-006322, PATH-006323, PATH-006324, PATH-006325, PATH-006326, PATH-006327, PATH-006328, PATH-006329, PATH-006332, PATH-006333, PATH-006344, PATH-006350, PATH-006351, PATH-006353, PATH-006382, PATH-006407, PATH-006409, PATH-006410, PATH-006447, PATH-006448, PATH-006451, PATH-006452, PATH-006471, PATH-006485, PATH-006486, PATH-006487, PATH-006488, PATH-006489, PATH-006490, PATH-006502, PATH-006505, PATH-006506, PATH-006522, PATH-006523, PATH-006524, PATH-006525, PATH-006526, PATH-006527, PATH-006528, PATH-006531, PATH-006533, PATH-006537, PATH-006538, PATH-006542, PATH-006543, PATH-006544, PATH-006545, PATH-006572, PATH-006573, PATH-006574, PATH-006602, PATH-006603, PATH-006604, PATH-006605, PATH-006606, PATH-006607, PATH-006608, PATH-006609, PATH-006610, PATH-006611, PATH-006612, PATH-006613, PATH-006614, PATH-006615, PATH-006630, PATH-006631, PATH-006632, PATH-006638, PATH-006650, PATH-006651, PATH-006652, PATH-006662, PATH-006663, PATH-006664, PATH-006665, PATH-006666, PATH-006667, PATH-006698, PATH-006712, PATH-006713, PATH-006714, PATH-006715, PATH-006716, PATH-006717, PATH-006735, PATH-006736, PATH-006738, PATH-006739, PATH-006740, PATH-006741, PATH-006742, PATH-006743, PATH-006744, PATH-006774, PATH-006830, PATH-006831, PATH-006853, PATH-006858, PATH-006859, PATH-006860, PATH-006861, PATH-006862, PATH-006872, PATH-006873, PATH-006874, PATH-006877, PATH-006878, PATH-006984. 4 additional serial numbers were added 6/5/12 - PATH-006332, PATH-006333, PATH-006350, and PATH-006351.
- 510(k) numbers
- ["K042745"]
- PMA numbers
- not on file
- Product code
- MNR
- Firm FEI
- 3002807992
- Firm city
- Hertford
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28