GE Medical Systems, LLC: SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System
- Recall number
- Z-0519-2025
- Recalling firm
- GE Medical Systems, LLC
- Product
- SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System
- Status
- Open, Classified
- Initiated
- 2024-10-28
- Posted
- 2024-11-22
- Terminated
- not on file
- Reason
- Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
- Root cause
- Device Design
- Action
- GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to recall.60998@gehealthcare.com. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 3 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
- Lot, serial or model codes
- UDI/DI (01)00195278283481(11)231214(21)CNARC2300080TJ, System ID 82427110161; UDI/DI (01)00195278283481(11)230519(21)CNARC2300027TJ, System ID 82427090471; UDI/DI (01)00195278283481(11)231116(21)CNARC2300072TJ, System ID 82427120353; UDI/DI (01)00195278283481(11)230830(21)CNARC2300051TJ, System ID 82427140303; UDI/DI (01)00195278283481(11)230515(21)CNARC2300026TJ, System ID 82427200210; UDI/DI (01)00195278283481(11)230809(21)CNARC2300046TJ, System ID 82427090487; UDI/DI (01)00195278283481(11)230816(21)CNARC2300048TJ, System ID 82427010214; UDI/DI (01)00195278283481(11)230719(21)CNARC2300040TJ, System ID 82427070331; UDI/DI (01)00195278283481(11)230904(21)CNARC2300052TJ, System ID 82427090477; UDI/DI (01)00195278283481(11)231102(21)CNARC2300068TJ, System ID 82427160409; UDI/DI (01)00195278283481(11)230722(21)CNARC2300041TJ, System ID 82427040320; UDI/DI (01)00195278283481(11)231120(21)CNARC2300073TJ, System ID 82427090470; UDI/DI (01)00195278283481(11)231009(21)CNARC2300060TJ, System ID 82427030219; UDI/DI (01)00195278283481(11)240104(21)CNARC2400002TJ, System ID 82427090488; UDI/DI (01)00195278283481(11)231207(21)CNARC2300078TJ, System ID 82427090469; UDI/DI (01)00195278283481(11)230918(21)CNARC2300056TJ, System ID 82427200214; UDI/DI (01)00195278283481(11)240131(21)CNARC2400008TJ, System ID 82427030223; UDI/DI (01)00195278283481(11)240113(21)CNARC2400005TJ, System ID 82427060120; UDI/DI (01)00195278283481(11)231109(21)CNARC2300070TJ, System ID 82427130176; UDI/DI (01)00195278283481(11)230920(21)CNARC2300057TJ, System ID 82427100480; UDI/DI (01)00195278283481(11)231010(21)CNARC2300061TJ, System ID 82427100491; UDI/DI (01)00195278283481(11)230729(21)CNARC2300043TJ, System ID 82427250136; UDI/DI (01)00195278283481(11)240116(21)CNARC2400006TJ, System ID 82427250132; UDI/DI (01)00195278283481(11)240110(21)CNARC2400004TJ, System ID 82427110160; UDI/DI (01)00195278283481(11)231026(21)CNARC2300066TJ, System ID 82427100496; UDI/DI (01)00195278283481(11)231204(21)CNARC2300077TJ, System ID 82427100504; UDI/DI (01)00195278283481(11)230508(21)CNARC2300024TJ, System ID 82427160403; UDI/DI (01)00195278283481(11)231221(21)CNARC2300081TJ, System ID 82427030207; UDI/DI (01)00195278283481(11)240102(21)CNARC2400001TJ, System ID 82427120364; UDI/DI (01)00195278283481(11)231019(21)CNARC2300064TJ, System ID 82427120360; UDI/DI (01)00195278283481(11)240127(21)CNARC2400007TJ, System ID 82427200224; UDI/DI (01)00195278283481(11)231030(21)CNARC2300067TJ, System ID 82427200207; UDI/DI (01)00195278283481(11)231016(21)CNARC2300063TJ, System ID 82427140310; UDI/DI (01)00195278283481(11)230818(21)CNARC2300049TJ, System ID 82427140317; UDI/DI (01)00195278283481(11)230911(21)CNARC2300054TJ, System ID 82427140308; UDI/DI (01)00195278283481(11)230913(21)CNARC2300055TJ, System ID 82427190195
- 510(k) numbers
- ["K160618","K163331","K202966"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 2183553
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28