GE Medical Systems, LLC: SIGNA Architect, Nuclear Magnetic Resonance Imaging System

Recall number
Z-0518-2025
Recalling firm
GE Medical Systems, LLC
Product
SIGNA Architect, Nuclear Magnetic Resonance Imaging System
Status
Open, Classified
Initiated
2024-10-28
Posted
2024-11-22
Terminated
not on file
Reason
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
Root cause
Device Design
Action
GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to recall.60998@gehealthcare.com. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity
53 units
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
Lot, serial or model codes
UDI/DI (01)00840682147095(11)240100(21)PG75M2400278SC, System ID 303467MR2; UDI/DI To be provided, System ID 1000005000; UDI/DI (01)00840682147095(11)220800(21)PG75M2300225SC, System ID 217528MR4; UDI/DI (01)00840682147095(11)220600(21)PG75M2200130SC, System ID 781SSHMR3; UDI/DI To be provided, System ID 301295AMR2; UDI/DI To be provided, System ID GON90303794; UDI/DI To be provided, System ID 215955HON3T; UDI/DI (01)00840682147095(11)230600(21)PG75M2300245SC, System ID 806UMCHWMR3T; UDI/DI (01)00840682147095(11)230800(21)PG75M2300255SC, System ID 703573ARC7; UDI/DI (01)00840682122702(11)220100(21)PG75A2300133SC, System ID 757953SPMR; UDI/DI (01)00840682147095(11)231000(21)PG75M2300263SC, System ID 905684MR3T; UDI/DI (01)00840682147095(11)220800(21)PG75M2200153SC, System ID 82427160355; UDI/DI To be provided, System ID 82427100482; UDI/DI To be provided, System ID 82427140330; UDI/DI To be provided, System ID 82427040333; UDI/DI To be provided, System ID 82427200209; UDI/DI To be provided, System ID 82427050061; UDI/DI To be provided, System ID 82427080138; UDI/DI (01)00840682147095(11)230400(21)PG75M2300232SC, System ID 82427150128; UDI/DI To be provided, System ID 82427200219; UDI/DI To be provided, System ID 82427080142; UDI/DI To be provided, System ID 82427230234; UDI/DI (01)00840682147095(11)231000(21)PG75M2300266SC, System ID 82427100499; UDI/DI To be provided, System ID 82427160405; UDI/DI To be provided, System ID 82427160407; UDI/DI (01)00840682147095(11)230600(21)PG75M2300247SC, System ID 82427160390; UDI/DI To be provided, System ID 82427160408; UDI/DI To be provided, System ID 82427160402; UDI/DI To be provided, System ID 82427120372; UDI/DI To be provided, System ID 82427120357; UDI/DI To be provided, System ID 82427120351; UDI/DI (01)00840682147095(11)230700(21)PG75M2300253SC, System ID 82427150130; UDI/DI (01)00840682147095(11)240100(21)PG75M2400276SC, System ID 82427030222; UDI/DI (01)00195278283481(11)231113(21)CNARC2300071TJ, System ID 82427030218; UDI/DI (01)00840682147095(11)230600(21)PG75M2300248SC, System ID 83027009451123; UDI/DI (01)00840682147095(11)240200(21)ANTAT2400005TJ, System ID MR00169; UDI/DI (01)00840682147095(11)240200(21)ANTAT2400004TJ, System ID MR00168; UDI/DI (01)00840682147095(11)240100(21)ANTAT2400001TJ, System ID MR00185; UDI/DI (01)00840682122702(11)180223(21)PG75A1800019MR, System ID 850270540; UDI/DI To be provided, System ID 850270791; UDI/DI To be provided, System ID LV4001MR03; UDI/DI (01)00840682147095(11)230600(21)PG75M2300246SC, System ID MRUA2408; UDI/DI To be provided, System ID P00434MR01; UDI/DI To be provided, System ID RF3138MR01; UDI/DI To be provided, System ID RU1838MR04; UDI/DI To be provided, System ID RF3580MR01; UDI/DI To be provided, System ID SA1058MR05; UDI/DI (01)00840682147095(11)221100(21)PG75M2200180SC, System ID FEG88320; UDI/DI (01)00840682147095(11)230800(21)PG75M2300256SC, System ID DE492A99; UDI/DI To be provided, System ID NEV439249; UDI/DI To be provided, System ID 853270212; UDI/DI To be provided, System ID 853270202; UDI/DI (01)00840682147095(11)230300(21)PG75M2300227SC, System ID 853270223
510(k) numbers
K160618, K163331, K202966
PMA numbers
not on file
Product code
LNH
Firm FEI
2183553
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source