GE Healthcare, LLC: GE Vscan Extend, being sold as "Vscan Extend Console"

Recall number
Z-0518-2022
Recalling firm
GE Healthcare, LLC
Product
GE Vscan Extend, being sold as "Vscan Extend Console"
Status
Open, Classified
Initiated
2021-11-04
Posted
2022-01-24
Terminated
not on file
Reason
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
Root cause
Device Design
Action
The firm distributed an Urgent Medical Device Correction letter to customers on 11/17/2021 informing customers of the issue with Vscan Extend devices not powering on because of depleted internal battery. This issue occurs if the device is not turned off prior to the external battery being removed or when the device is connected to a personal computer without connecting to AC power. The actions needed to be taken by the user until a software update could be deployed were identified in this notice as: make sure to power off the Vscan Extend device before removing the external battery, always connect the device to AC power when it is connected to a PC, and if the device is being used outside of a normal hospital or clinic setting where it is the only imaging device available, ensure the device functionality is checked daily or before each shift. Attached to the notification was an Acknowledgment Response form that was to be returned by email within 30 days of receipt of the notice. Customers with any questions are to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
Quantity
12,526 total reported units (7 units of this model)
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR, and DC. International: Albania, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Ecuador, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Haiti, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Moldova, Morocco, Myanmar, Netherlands, New Zealand, Norway, Papua New Guinea, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Somalia, South Africa, Spain, Sweden, Switzerland, Thailand, Togo, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam, and Western Sahara.
Lot, serial or model codes
Information reads "Serial Number (UDI, Model Number)". VH1622261Y (01008406821216511118042621VH1622261Y, 5693455), VH161664XA (01008406821216511118031921VH161664XA, 5693455), VH2627194S (01008406821216511120071321VH2627194S, 5693455), VH160129E8 (01008406821216511117021721VH160129E8, 5693455), VH160834YQ (01008406821216511117110021VH160834YQ, 5693455), VH1642124F (01008406821216511119040921VH1642124F, 5693455), VH150152W7 (01008406821216511116080021VH150152W7, 5693455), VH160113JQ (01008406821216511117020321VH160113JQ, 5693455), VH161979EY (01008406821216511118030021VH161979EY, 5693455), VH162446PC (01008406821216511118061221VH162446PC, 5693455), VH161984GS (01008406821216511118030021VH161984GS, 5693455), VH161970GQ (01008406821216511118030021VH161970G, 5693455), VH1619899S (01008406821216511118030021VH1619899S, 5693455), VH160237FQ (01008406821216511117030021VH160237FQ, 5693455), VH1600005U (5693455), VH160007CX (5693455), VH160009UF (5693455), VH150153FK (5693455), VH161672SF (5693455), VH1624281N (5693455), VH150103AI (5693455), VH150150Z6 (5693455), VH160786JM (5693455), VH1608086P (5693455), VH1600107T (5693455), VH1607384V (5693455), VH1600112G (5693455), VH160012JB (5693455), VH1607406X (5693455), VH160742EX (5693455), VH1608215J (5693455), VH160633NC (5693455), VH160025HU (5693455), VH160744ZM (5693455), VH150151QA (5693455), VH160730NB (5693455), VH161629J2 (5693455), VH161631UU (5693455), VH160735W7 (5693455), VH161598QQ (5693455), VH1600189D (5693455), VH1501544T (5693455), VH160017E2 (5693455), VH150123AH (5693455), VH163240NI (5693455), VH160005HB (5693455), VH1656702Z (5693455), VH16567191 (5693455), VH1656729W (5693455), VH1656736Y (5693455), VH165674RL (5693455), VH165675AQ (5693455), VH165676D7 (5693455), VH1656777R (5693455), VH165678YC (5693455), VH165679H1 (5693455), VH150112VD (5693455), VH150113IF (5693455), VH1607343V (5693455), VH160012JB (5693455), VH150101S2 (5693455), VH160001S1 (5693455), VH160003N9 (5693455), VH1616079K (5693455), VH150105RI (5693455), VH150118BJ (5693455), VH1501144B (5693455), VH150119KV (5693455), VH150122PV (5693455), VH150104JE (5693455), VH150067PS (5693455), VH150102JE (5693455), VH16162756 (5693455), VH260391G7 (5693455), VH2609418F (5693455), VH26096618 (5693455), VH1600301G (5693455), VH160021FJ (5693455), VH160163HW (5693455)
510(k) numbers
["K161588","K180995"]
PMA numbers
not on file
Product code
IYO
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register