Laerdal Medical Corporation: LAERDAL Silicone Resuscitator (LSR), Pediatric Size, containing an Intake/Reservoir Valve. Sold as Part Nos: 860050-Pediatric Basic w/o Mask in Carton; 860051-Pediatric Complete in Carton; 860052-Standard Child w/Child Mask 3-4 in Carton; 860053-Pediatric Complete in Compact Case; 860055-Pediatric Complete in Display Case; 860056-Standard Term-Baby w/Infant Mask 0-1 in Carton; 865500-Pediatric Upgrade Kit . Responsible firm on the label: Distributed in the U.S. by Laerdal Medical Corporation, Wappingers Falls, NY, Made in Norway.

Recall number
Z-0518-04
Recalling firm
Laerdal Medical Corporation
Product
LAERDAL Silicone Resuscitator (LSR), Pediatric Size, containing an Intake/Reservoir Valve. Sold as Part Nos: 860050-Pediatric Basic w/o Mask in Carton; 860051-Pediatric Complete in Carton; 860052-Standard Child w/Child Mask 3-4 in Carton; 860053-Pediatric Complete in Compact Case; 860055-Pediatric Complete in Display Case; 860056-Standard Term-Baby w/Infant Mask 0-1 in Carton; 865500-Pediatric Upgrade Kit . Responsible firm on the label: Distributed in the U.S. by Laerdal Medical Corporation, Wappingers Falls, NY, Made in Norway.
Status
Terminated
Initiated
2003-12-23
Posted
2004-02-10
Terminated
2005-11-16
Reason
Potential for the Flap Valve component of the Intake/Reservoir Valve to become dislodged from its mount, rendering the resuscitator inoperative.
Root cause
Other
Action
Letters issued on 1/14/2004. The recall involves the replacement of the Inner Part of the Intake/Reservoir Valve, which is part of the Laerdal Silicone Resuscitator. Recall is to the user level.
Quantity
766 units
Distribution
Product was shipped to approximately 264 customers nationwide, and 3 foreign consignees (Mexico, Brazil and Northern Mariana Islands). Product was also sold to a few exporters who shipped units to unknown countries. Customers include distributors and end users. There were 2,403 parts shipped to Laerdal Medical Canada, Ltd., Toronto, Ontario, Canada. That firm is responsible for conducting an independent recall, reportedly with the knowledge of the Canadian government.
Lot, serial or model codes
All product shipped after 1/29/2003 is affected. The following lot numbers are the lot numbers for adult, pediatric and preterm size LSR: 0403, 0703, 0803, 0903, 1003, 1103, 1203, 1303, 1403, 1503, 1603, 1703, 1803, 1903, 2003, 2103, 2203, 2303, 2503, 2703, 2903, 3103, 3203, 3303, 3503, 3603, 3703, 3803, 3903, 4003, 4103, 4203, 4303, 4403, 4503, 4603, 4703, 4803, 4903. NOTE: customers were NOT provided with lot numbers. They were instructed to visually examine the Intake/Reservoir valve in order to determine if they have affected product.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BTM
Firm FEI
2425852
Firm city
Wappingers Falls
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register