Siemens Healthcare Diagnostics, Inc.: Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension(R) clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of Tacrolimus therapy in kidney and liver transplant patients

Recall number
Z-0517-2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Dimension(R) TACR Flex(R) reagent cartridge (DF107) The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension(R) clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of Tacrolimus therapy in kidney and liver transplant patients
Status
Terminated
Initiated
2013-04-25
Posted
2013-12-16
Terminated
2014-11-12
Reason
Siemens has confirmed that the TACR method may demonstrate reduced on-board stability which may result in imprecise and inaccurate QC and patient results.
Root cause
Component design/selection
Action
Siemens sent an Urgent Medical Device Correction letter dated April 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. A recall notification letter was sent to consignees in April 2013. Customers using the affected product should load Flex reagent cartridges on board immediately prior to use and remove the Flex after 8 hours. Do not load more Flexes than the lab will use in 8 hours and do not pre-hydrate TACR Flexes. Customers were asked to complete the attached Effectiveness Check form and fax it to the Siemens Technical Solutions Center at 302-631-8467. Customers were asked to forward the communication to anyone to whom may have distributed the product. Customers with questions were instructed to contact the Siemens Technical Solutions Center or their local Siemens technial support representative. For questions regarding this recall call 800-441-9250.
Quantity
13,751
Distribution
Nationwide Distribution including AK, AL, AR, AZ, CA, DC, FL, GA, IA, IL, LA, MD, MI, MO, MS, NC, NH, NJ, NM, OH, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, and WI.
Lot, serial or model codes
Siemens Material Number 10444938, lot numbers GB3099, GA3120, DB3141, GB3162, GB3176, FA3197, FA3267 and FB3267.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MLM
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register