Siemens Medical Solutions USA, Inc: syngo Imaging XS, versions VA70A and VA70B Client/Server Installations Radiological image processing system

Recall number
Z-0515-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
syngo Imaging XS, versions VA70A and VA70B Client/Server Installations Radiological image processing system
Status
Terminated
Initiated
2011-09-19
Posted
2012-01-11
Terminated
2014-01-24
Reason
Unintended behavior when using syngo Imaging XS, versions VA70A and VA70B Client/Serve installations. Loading additional images/series from Patient Browser using the thumbnails can cause an unintended deletion of images/series.
Root cause
Software design
Action
Siemens Medical Solutions USA, Inc sent a Customer Safety Advisory letter dated August 24, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you detect patient data inconsistenceis in your system caused by this problem, please contact Siemens Service for support. For further questions please call (610) 219-4834.
Quantity
12
Distribution
Nationwide Distribution including the states of CA, IN, IA, NV, NE, NY, OH, TN and TX.
Lot, serial or model codes
Model number 10496279
510(k) numbers
["K082430"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register