Straumann USA LLC: Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089

Recall number
Z-0512-2022
Recalling firm
Straumann USA LLC
Product
Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089
Status
Terminated
Initiated
2021-11-30
Posted
not on file
Terminated
2025-02-05
Reason
Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm implant
Root cause
Packaging change control
Action
Neodent issued Medical Device Recall letter initial contact by phone initiated on November 30th, 2021; followed by Confirmation written letter via UPS December 3rd 2021. Letter states reason for recall, health risk and action to take: Search in your stock for units according to the description on the table above. " If the implant has not yet been used, identify and segregate the affected units, open or closed, in your stock. Return the product using the label included in this mailing, refer to Field Action . Upon receipt of the product by Straumann, replacement product will be sent. If the product is not found in your inventory, indicate on the Form below. " If the product has been implanted and no issue was identified, it is not necessary to remove, additional patient follow-up is not required. Indicate this on the Customer Confirmation Form (included) " If the product has been installed and an issue has been reported by the patient, for example: numbness, pain or any other symptom, please communicate immediately to Straumann. " For either cases, complete and return the enclosed Customer Confirmation Form to Action 003/2021. 3.1. DATA FOR PRODUCT RETURN: Straumann USA Regulatory Affairs Action 003/2021 Product Return 60 Minuteman Road, Andover, MA 01810
Quantity
7 pieces
Distribution
US Nationwide Distribution in the states of FL, IL, OH
Lot, serial or model codes
Lot GNV17 UDI: 7899878052722
510(k) numbers
["K203382"]
PMA numbers
not on file
Product code
DZE
Firm FEI
1000121052
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register