Skip to the content

FDA Medical Device Recalls (openFDA)

Teco Diagnostics: Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise Listing number: D072377 Each device consists of 100 strips of Urine Reagent Strips 10 parameters packed with a desiccant pouch inside a gray bottle with Teco Diagnostics Label on the bottle and the bottle placed inside a Teco Diagnostics Box. For the semi-quantitative and qualitative detection of Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes in urine.

Recall number
Z-0511-2012
Recalling firm
Teco Diagnostics
Product
Urine Reagent Strips (URS) 10 Parameters URS-l0-100W 510(K) number: K970250 Devise Listing number: D072377 Each device consists of 100 strips of Urine Reagent Strips 10 parameters packed with a desiccant pouch inside a gray bottle with Teco Diagnostics Label on the bottle and the bottle placed inside a Teco Diagnostics Box. For the semi-quantitative and qualitative detection of Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes in urine.
Status
Terminated
Initiated
2011-10-14
Posted
2012-01-11
Terminated
2013-04-09
Reason
This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014. Discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets (Refer to next paragraph). Sampling size submitted for QC testing was insuff
Root cause
Nonconforming Material/Component
Action
Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
Quantity
550 kits
Distribution
Distributed in CA and Poland.
Lot, serial or model codes
Product/Catalog # 1: URS-l 0-100W i. Lot Number: 35264 ii. Expiration Date: 01/2014
510(k) numbers
not on file
PMA numbers
not on file
Product code
KQO
Firm FEI
1832216
Firm city
Anaheim
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register