Applied Medical Resources Corp: GelPOINT Path Transanal Procedural Pack with Epix Electrosurgical L-Hook with Smoke Evacuation, 4x5.5cm and 5mm x 42cm REF TA211 Product Usage: The GelPOINT Path Transanal Access Platform is indicated for multiple instrument or camera access through the anus to perform various diagnostic and/or therapeutic procedures by using additional instruments.
- Recall number
- Z-0510-2020
- Recalling firm
- Applied Medical Resources Corp
- Product
- GelPOINT Path Transanal Procedural Pack with Epix Electrosurgical L-Hook with Smoke Evacuation, 4x5.5cm and 5mm x 42cm REF TA211 Product Usage: The GelPOINT Path Transanal Access Platform is indicated for multiple instrument or camera access through the anus to perform various diagnostic and/or therapeutic procedures by using additional instruments.
- Status
- Terminated
- Initiated
- 2018-12-12
- Posted
- not on file
- Terminated
- 2021-01-22
- Reason
- Unintentional shifting of the tip insulation of the firm's electrosurgical probes.
- Root cause
- Device Design
- Action
- On12/12/2018, the firm sent an "URGENT; MEDICAL DEVICE RECALL" notification to customers via UPS informing them that this voluntary recall is being conducted due to the possibility that the insulation near the device tip may unintentionally shift. Customers are asked to: Check their inventory for recalled product. Complete the attached Recall Notification Confirmation Form to acknowledge the recall and indicate if their facility is returning or has already used units or kits from the lots listed. o If no product is being returned, to please indicate that on the Recall Notification Confirmation Form. Provide a no-charge P.O. number if replacement units or kits are requested. If customer is a distributor, they are instructed to notify any facilities to which they distributed units or kits from the affected lots, and to have them complete the appropriate sections of the Recall Notification Confirmation Form. Return the completed Recall Notification Confirmation Form to Applied Medical by emailing it to recall60787748@appliedmedical.com or faxing it to 949-713-8871. Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical following the instructions provided. For product return questions, please contact the Manager of Customer Relations, RGA Dept. at (949) 713-8622 or by email at kmitchell@appliedmedical.com. For regulatory questions, to contact (949) 713-8767 or by email at lcontursi@appliedmedical.com.
- Quantity
- 39 units Domestically and 112 units Internationally
- Distribution
- Worldwide distribution - US Nationwide distribution and countries of Belgium; Germany; Switzerland; Spain; Finland; France; Great Britain; Israel; India; Italy; Lebanon; Netherlands; and Portugal.
- Lot, serial or model codes
- Model Number: TA211 Lot Numbers: 1342169 and 1340118
- 510(k) numbers
- ["K171701"]
- PMA numbers
- not on file
- Product code
- FER
- Firm FEI
- 2027111
- Firm city
- Rancho Santa Margarita
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28