FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
- Recall number
- Z-0509-2026
- Recalling firm
- Medtronic Perfusion Systems
- Product
- MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
- Status
- Open, Classified
- Initiated
- 2025-10-03
- Posted
- 2025-11-14
- Terminated
- not on file
- Reason
- As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
- Root cause
- Software in the Use Environment
- Action
- Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Representative.
- Quantity
- 167 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- UDI/DI 10850007732247, Serial Numbers: 000023, 000047, 000073, 000100, 000126, 000150, 000173, 000197, 000223, 000024, 000048, 000075, 000101, 000127, 000151, 000174, 000202, 000224, 000025, 000051, 000076, 000102, 000128, 000152, 000177, 000203, 000225, 000027, 000052, 000078, 000103, 000129, 000153, 000179, 000204, 000228, 000028, 000054, 000079, 000104, 000130, 000154, 000180, 000205, 000229, 000029, 000055, 000080, 000106, 000131, 000155, 000181, 000206, 000277, 000030, 000056, 000081, 000108, 000132, 000156, 000182, 000207, 000284, 000031, 000058, 000082, 000109, 000133, 000157, 000183, 000208, 000032, 000059, 000083, 000110, 000134, 000159, 000184, 000209, 000033, 000060, 000084, 000112, 000135, 000161, 000185, 000210, 000034, 000061, 000085, 000115, 000136, 000162, 000186, 000211, 000035, 000063, 000087, 000116, 000140, 000163, 000187, 000212, 000037, 000065, 000088, 000117, 000141, 000165, 000188, 000213, 000038, 000066, 000089, 000118, 000142, 000166, 000189, 000214, 000039, 000067, 000090, 000119, 000143, 000167, 000190, 000215, 000040, 000068, 000092, 000120, 000145, 000168, 000191, 000216, 000043, 000069, 000095, 000121, 000146, 000169, 000192, 000217, 000044, 000070, 000096, 000122, 000147, 000170, 000193, 000218, 000045, 000071, 000097, 000124, 000148, 000171, 000194, 000219, 000046, 000072, 000099, 000125, 000149, 000172, 000195, 000221.
- 510(k) numbers
- ["K230364","K250199"]
- PMA numbers
- not on file
- Product code
- QNR
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28