FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Recall number
- Z-0508-2018
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- AArtis Q/ Q.Zen biplane Artis zee biplane Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Status
- Terminated
- Initiated
- 2017-08-24
- Posted
- not on file
- Terminated
- 2018-08-29
- Reason
- Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
- Root cause
- Employee error
- Action
- Siemens sent an Urgent: Medical Device Correction letter dated August 24, 2017 to all affected customers. The letter identified the affected product, problem and actions to be taken. A local management or dispatch will schedule for an inspection to correct the issue. For questions please call 1-800-888-7436
- Quantity
- 265 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- 10094141 10848282 10848355
- 510(k) numbers
- ["K073290","K123529"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28