Skip to the content

FDA Medical Device Recalls (openFDA)

Alphatec Spine, Inc.: Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.

Recall number
Z-0507-2009
Recalling firm
Alphatec Spine, Inc.
Product
Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.
Status
Terminated
Initiated
2007-08-16
Posted
2008-12-31
Terminated
2009-01-12
Reason
Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were mixed from the following lot numbers: Part Number 64733-120 Lot Number 611884, and Part Number 64753-126 and Lot Number 611881.
Root cause
Process control
Action
The recall was initiated on August 16, 2007 with Alphatec Spine contacting each of the four direct distributors by telephone. This was followed with a Product Recall Notice, dated August 27, 2007, that was mailed to all affected customers. The distributors were informed that the firm was conducting a voluntary recall because two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. The firm asked the customers to return the product, and were informed that replacement product will be sent back.
Quantity
3 units
Distribution
Nationwide Distribution --- including states of AZ, FL, NY & TX.
Lot, serial or model codes
Lot Number: 611884
510(k) numbers
not on file
PMA numbers
not on file
Product code
MQP
Firm FEI
2027467
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register