FDA Medical Device Recalls (openFDA)
Alphatec Spine, Inc.: Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.
- Recall number
- Z-0507-2009
- Recalling firm
- Alphatec Spine, Inc.
- Product
- Novel PEEK VBR XS 16-22 MM, S-5, 20MM, Part Number: 64733-120. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.
- Status
- Terminated
- Initiated
- 2007-08-16
- Posted
- 2008-12-31
- Terminated
- 2009-01-12
- Reason
- Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were mixed from the following lot numbers: Part Number 64733-120 Lot Number 611884, and Part Number 64753-126 and Lot Number 611881.
- Root cause
- Process control
- Action
- The recall was initiated on August 16, 2007 with Alphatec Spine contacting each of the four direct distributors by telephone. This was followed with a Product Recall Notice, dated August 27, 2007, that was mailed to all affected customers. The distributors were informed that the firm was conducting a voluntary recall because two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. The firm asked the customers to return the product, and were informed that replacement product will be sent back.
- Quantity
- 3 units
- Distribution
- Nationwide Distribution --- including states of AZ, FL, NY & TX.
- Lot, serial or model codes
- Lot Number: 611884
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- MQP
- Firm FEI
- 2027467
- Firm city
- Carlsbad
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28