Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Pneumothorax Tray, Catalog No. C-UTPTY-1400-WAYNE-112497-IMH, GPN G56537 - Product Usage: The modified Wayne Pneumothorax Set is used for the relief of simple, spontaneous, iatrogenic, and tension pneumothorax.

Recall number
Z-0506-2020
Recalling firm
Cook Inc.
Product
Pneumothorax Tray, Catalog No. C-UTPTY-1400-WAYNE-112497-IMH, GPN G56537 - Product Usage: The modified Wayne Pneumothorax Set is used for the relief of simple, spontaneous, iatrogenic, and tension pneumothorax.
Status
Terminated
Initiated
2019-09-17
Posted
not on file
Terminated
2020-05-21
Reason
Lidstock perforation line may be over the sterile area of the tray, when it should be on the lip of the tray, potentially compromising the sterility of the device.
Root cause
Under Investigation by firm
Action
On September 17, 2019, the firm notified consignees of the recall by mailing Urgent Medical Device Recall letters. The letter alerted customers that if the affected product is used, they may experience increased procedural time to obtain a replacement and local or systemic infection. Customers were asked to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com.
Quantity
41 units
Distribution
US Nationwide distribution in the states of CA, FL, IN, MI, MT, NC, NY, OH.
Lot, serial or model codes
Lot 9850522
510(k) numbers
not on file
PMA numbers
not on file
Product code
JOL
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register