Siemens Healthcare Diagnostics, Inc: Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 100 Test 02816138 10309045, 10309044 028 Intended Use For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure.

Recall number
Z-0506-2013
Recalling firm
Siemens Healthcare Diagnostics, Inc
Product
Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 100 Test 02816138 10309045, 10309044 028 Intended Use For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure.
Status
Terminated
Initiated
2012-11-14
Posted
2012-12-11
Terminated
2015-09-25
Reason
Operating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.
Root cause
Device Design
Action
Siemens Healthcare issued An Urgent Medical Device Correction Notice (for US customers) and an Urgent Field Safety Notice (for customers outside the US) dated November 5, 2012, to all affected customers. These notices inform customers who used the ADVIA Centaur BNP, BR, CA 19-9, Digitoxin, Folate, and Theophylline 2 assays to reduce their instrument operating temperature range to 20 -25 degrees C and for the DHEAS assay to reduce the instrument operating temperature range to 18 -26 degrees C when these assays are being used. These notices also direct customers to review the contents of the correction notice with their Laboratory Director or Medical Director. A confirmation fax-back form was included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local technical support representative. For questions regarding this recall call 508-668-5000.
Quantity
100 test=24291 and 500 test= 9679
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam
Lot, serial or model codes
All lot codes
510(k) numbers
["K031038"]
PMA numbers
not on file
Product code
NBC
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register