FDA Medical Device Recalls (openFDA)
Philips North America Llc: Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.
- Recall number
- Z-0505-2024
- Recalling firm
- Philips North America Llc
- Product
- Panorama 1.0T HFO, Magnetic Resonance (MR) System; Model Numbers 781250 and 781350. Philips Magnetic Resonance (MR) systems are Medical Electrical Systems for use as a diagnostic device.
- Status
- Open, Classified
- Initiated
- 2023-11-17
- Posted
- 2023-12-14
- Terminated
- not on file
- Reason
- Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.
- Root cause
- Component design/selection
- Action
- On November 17, 2023, the firm issued an URGENT Medical Device Correction letter to affected consignees. Customers were provided instructions on how to identify impacted systems. Customers with impacted systems should immediately discontinue use of their impacted MR system. Philips is requesting that customers immediately stop using the MR systems until an inspection can be completed. Philips will contact customers to schedule an on-site visit by a Field Service Engineer to inspect your system starting in December 2023. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity
- 150 US; 192 ROW
- Distribution
- Domestic distribution to AL AZ CA CO DE FL GA IA IL IN KS KY LA MA MD MI MO NE NJ NV NY OH OK OR PA SC TN TX UT VA VT WA WI WV International distribution to Argentina Austria Canada China Colombia Croatia Czech Republic Denmark Ecuador Egypt France Germany Greece Italy Kuwait Libya Netherlands NO93 Panama Poland Portugal Qatar Russian Fed. Saint Kitts Saudi Arabia Spain Sweden Switzerland Syria Thailand T¿rkiye United Kingdom Utd.Arab.Emir. Venezuela
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K041602","K052078","K063559"]
- PMA numbers
- not on file
- Product code
- LNH
- Firm FEI
- 3016618143
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28