FDA Medical Device Recalls (openFDA)
Becton Dickinson & Company: BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;
- Recall number
- Z-0504-2026
- Recalling firm
- Becton Dickinson & Company
- Product
- BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;
- Status
- Open, Classified
- Initiated
- 2025-10-14
- Posted
- 2025-11-14
- Terminated
- not on file
- Reason
- BD has confirmed through internal testing that certain device trays failed routine biocompatibility testing.
- Root cause
- Under Investigation by firm
- Action
- On October 14, 2025, URGENT: MEDICAL DEVICE REMOVAL letters were sent to distributors and customers. Please Take the Following Actions: 1. Cease use and destroy any unused affected product according to your facilities regulations. BD will provide credit for all product destroyed. 2. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on BD s behalf. 3. Complete the attached Distributor Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 5. Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com
- Quantity
- 4180000 units
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, Singapore, Hong Kong.
- Lot, serial or model codes
- Catalog Number(s) or Model Number(s): 308341; UDI-DI: 50382903083416; Lot Numbers: 5029670, 5034211, 5041975;
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KYW
- Firm FEI
- 2243072
- Firm city
- Franklin Lakes
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28